51
pH
Memosens pH input
(also ISFET)
M8 socket, 4 pins, for Memosens lab cable or
M12 socket, 8 pins, for Memosens sensors
Display ranges
1)
pH
-2.00 … +16.00
mV
-1999 … +1999 mV
Temperature
-50 … +250 °C (-58 ... +482 °F)
Memosens input
ORP
M8 socket, 4 pins, for Memosens lab cable or
M12 socket, 8 pins, for Memosens sensors
Display ranges
1)
mV
-1999 … +1999 mV
Temperature
-50 … +250 °C (-58 ... +482 °F)
Sensor standardization *
ORP calibration (zero adjustment)
Permissible calibration range
ΔmV (offset)
-700 … +700 mV
Sensor standardization
*
pH calibration
Operating modes *
PROG.BUFFERS
Calibration with automatic buffer recognition
MANUAL
Manual calibration with entry of individual
buffer values
DATA INPUT
Data entry of zero and slope
-01- Hauser
2.00/4.01/6.98/9.95/11.87
“Programmed buffers”
buffer sets *
-02- Mettler-Toledo
2.00/4.01/7.00/9.21
-03- Knick CaliMat
2.00/4.00/7.00/9.00/12.00
-04- Ciba (94)
2.06/4.00/7.00/10.00
-05- NIST technical
1.68/4.00/7.00/10.01/12.46
-06- NIST standard
1.679/4.006/6.865/9.180
-07- HACH
4.01/7.00/10.01/12.00
-08- WTW techn. buffers 2.00/4.01/7.00/10.00
-09- Hamilton
2.00/4.01/7.00/10.01/12.00
-10- Reagecon
2.00/4.00/7.00/9.00/12.00
-11- DIN 19267
1.09/4.65/6.79/9.23/12.75
Permissible calibration range
Zero point
6 ... 8 pH
With ISFET:
Operating point
(asymmetry)
-750 … +750 mV
Slope
Approx. 74 … 104 %
(possibly restricting notes from Sensoface)
Calibration timer
*
Interval 1 … 99 days, can be switched off
Sensoface
Provides information on the sensor condition
Evaluation of
zero/slope, response, calibration interval
* User-defined
1) Ranges depending on Memosens sensor
Specifications