29
Oxy
Oxygen Calibration
DATA INPUT Calibration
(Calibration by entering known sensor values)
The calibration method is selected in the configuration menu.
Measurement
CAL
DATA INPUT
xx blinks
nA
ZERO POINT
Use
st
to adjust the known value for the
sensor zero point.
xxx blinks
nA
SLOPE
Use
st
to adjust the known value for the
sensor slope.
Calibration will be performed. Automatic return to measuring mode.
Please note: To abort calibration, you can press
at any time.
Oxygen Calibration