Installation
Liquicap M FTI51, FTI52
26
Hauser
3.2
Installation instructions
Installation conditions
Liquicap M FTI51 (rod probe) can be installed from above, from below and from the side.
Liquicap M FTI52 (rope probe) can be installed vertically from above.
!
Note!
• The probe may not come into contact with the container wall! Do not install probes in the area
of the filling curtain!
• When using in agitating tanks, make sure you install at a safe distance from the agitator.
• If there are strong lateral loads, use rod probes with ground tube.
For electrically conductive tanks e.g. steel tanks
L00-FTI5xxxx-11-06-xx-xx-001
For nonconductive tanks e.g. plastic tanks
L00-FTI5xxxx-11-06-xx-xx-002
Probes with ground tube and grounding