background image

Levelflex FMP53 FOUNDATION Fieldbus

Electrical connection

Hauser

11

Terminal assignment PROFIBUS PA / FOUNDATION Fieldbus

A

1

+

2

-

3

+

4

-

B

1

+

2

-

3

+

4

-

1

3

5

5

2

4

  A0036500

 2

Terminal assignment PROFIBUS PA / FOUNDATION Fieldbus

A

Without integrated overvoltage protection

B

With integrated overvoltage protection

1

Connection, PROFIBUS PA / FOUNDATION Fieldbus: terminals 1 and 2, without integrated
overvoltage protection

2

Connection, switch output (open collector): terminals 3 and 4, without integrated overvoltage
protection

3

Connection, switch output (open collector): terminals 3 and 4, with integrated overvoltage protection

4

Connection, PROFIBUS PA / FOUNDATION Fieldbus: terminals 1 and 2, with integrated overvoltage
protection

5

Terminal for cable shield

Block diagram PROFIBUS PA / FOUNDATION Fieldbus

+

+

-

-

Y

I

2

3

1

2

3

4

1

4

  A0036530

 3

Block diagram PROFIBUS PA / FOUNDATION Fieldbus

1

Cable screen; observe cable specification

2

Connection PROFIBUS PA / FOUNDATION Fieldbus

3

Measuring device

4

Switch output (open collector)

6.1.2  

Device plug

In the case of the device versions with a plug, the housing does not need to be opened to

connect the signal cable.

Summary of Contents for Levelflex FMP53 FOUNDATION Fieldbus

Page 1: ...s they are not a substitute for the Operating Instructions pertaining to the device Detailed information about the device can be found in the Operating Instructions and the other documentation Availab...

Page 2: ...ord cd Ser no www endress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555 2 About this document 2 1 Symbols 2 1 1 Safety symbols DANGE...

Page 3: ...ls that must be connected to ground prior to establishing any other connections The ground terminals are located on the inside and outside of the device Inner ground terminal protective earth is conne...

Page 4: ...uthorized by the plant owner operator Are familiar with federal national regulations Must have read and understood the instructions in the manual and supplementary documentation Follow instructions an...

Page 5: ...with surfaces In the event of high medium temperatures ensure protection against contact to prevent burns 3 3 Workplace safety When working on and with the device Wear the required personal protectiv...

Page 6: ...onment the degree of protection indicated on the nameplate is no longer valid This may also impair the safe operation of the device 3 5 1 CE mark The measuring system meets the legal requirements of t...

Page 7: ...f the information on the measuring device and on the scope of the technical documentation pertaining to the device is displayed 4 3 Storage and transport 4 3 1 Storage temperature Permitted storage te...

Page 8: ...ents when mounting Distance A between the vessel wall and rod probes For smooth metallic walls 50 mm 2 in For plastic walls 300 mm 12 in to metallic parts outside the vessel Distance B between rod pro...

Page 9: ...curing screw using an open ended wrench 2 Rotate the housing in the desired direction 3 Tighten the securing screw 1 5 Nm for plastic housing 2 5 Nm for aluminum or stainless steel housing 5 2 2 Turni...

Page 10: ...en the housing and main electronics module and plug the display module into the electronics compartment until it engages Closing the cover of the electronics compartment 1 2 2 5 Nm 3 mm A0021451 1 Scr...

Page 11: ...nd 4 without integrated overvoltage protection 3 Connection switch output open collector terminals 3 and 4 with integrated overvoltage protection 4 Connection PROFIBUS PA FOUNDATION Fieldbus terminals...

Page 12: ...ge E 2 wire FOUNDATION Fieldbus switch output G 2 wire PROFIBUS PA switch output Non hazardous Ex nA Ex nA ia Ex ic Ex ic ia Ex d ia XP Ex ta DIP CSA GP 9 to 32 V 3 Ex ia IS Ex ia Ex d ia IS XP 9 to 3...

Page 13: ...Connect the potential matching line to the outer ground terminal before applying the power supply Required tools accessories For devices with a cover lock Allen key AF3 Wire stripper When using strand...

Page 14: ...sing shielded cables Connect the cable shield to the ground terminal 6 2 3 Plug in spring force terminals The electrical connection of device versions without an integrated overvoltage protection is v...

Page 15: ...compartment 2 Turn the securing clamp 90 in the clockwise direction and using an Allen key 3 mm tighten the screw of the securing clamp on the connection compartment cover with 2 5 Nm 7 Operation opt...

Page 16: ...or symbol if an error is active 1 2 Measured value symbols 1 3 Measured value 1 4 Unit 2 Measured value display bar graph 1 value 2 1 Bar graph for measured value 1 2 2 Measured value 1 including unit...

Page 17: ...8 Display module 1 Operating keys Key assignment Key Navigate down in the picklist Edit the numerical values or characters within a function Key Navigate up in the picklist Edit the numerical values...

Page 18: ...rement should generally be completely configured Setup Mapping Mapping of interference echos Setup Advanced setup Contains additional submenus and parameters For more customized configuration of the m...

Page 19: ...l display appears Calling up the menu via the context menu 1 Open the context menu 2 Press to navigate to the desired menu 3 Press to confirm the selection The selected menu opens 8 Commissioning 8 1...

Page 20: ...0 Operation Setup Main menu 0104 1 Display language English Espa ol Fran ais Display language English Deutsch 0104 1 Espa ol Fran ais Display language English Deutsch 0104 1 Betrieb Setup Hauptmen Spr...

Page 21: ...t D Distance E Empty calibration zero point L Level F Full calibration span 1 Setup Device tag Enter device tag 2 Setup Distance unit Select the length unit 3 Setup Tank type Select tank type 4 Setup...

Page 22: ...easured level L for verification purposes 9 Setup Distance Displays the distance D between the reference point R and the level L for verification purposes 10 Setup Signal quality Displays the signal q...

Page 23: ......

Page 24: ...www addresses endress com 71565960 71565960...

Reviews: