Diagnostics and troubleshooting
iTEMP TMT85
42
Hauser
Category
No.
Status messages
•
ACTUAL_STAT
US_NUMBER
in the
'Advanced
Diagnostics'
Transducer
Block
•
Local display
Error messages in the
relevant Sensor
Transducer Block
Sensor
Transducer Block
measured value
status (default)
Cause of error /
remedy
Output variables
affected
M-
103
Device status
message (FF):
Sensor drift/
difference
M-103
Local display:
M103 ↔
Measured value
BLOCK_ERR = Other
QUALITY =
UNCERTAIN
(configurable)
Cause of error:
Sensor drift has
been detected (in
accordance with
the settings in the
Advanced
Diagnostics Block).
Remedy:
Check the sensor,
depending on the
application.
PV1, PV2 SV1,
SV2
Transducer_Error = No
error
SUBSTATUS = Non
specific
M-
104
Device status
message (FF):
Backup active
M-104
Local display:
M104 ↔
Measured value
BLOCK_ERR = Other
QUALITY =
GOOD / BAD
Cause of error:
Backup function
activated and an
error was detected
at a sensor.
Remedy:
Resolve sensor
error.
SV1, SV2, also
PV1, PV2
depending on the
configuration
Transducer_Error = No
error
SUBSTATUS = Non
specific
9.2.3
Category S diagnostics messages
Category
No.
Status messages
•
ACTUAL_STAT
US_NUMBER
in the
'Advanced
Diagnostics'
Transducer
Block
•
Local display
Error messages in the
relevant Sensor
Transducer Block
Sensor
Transducer Block
measured value
status (default)
Cause of error /
remedy
Output variables
affected
S-
502
Device status
message (FF):
Special
linearization
S-501
Local display:
S501 ↔ Measured
value
BLOCK_ERR = Other |
Block Configuration
Error
QUALITY = BAD
Cause of error:
Linearization
error.
Remedy:
Select valid type of
linearization
(sensor type).
SV1, SV2, PV1,
PV2, DT
Transducer_Error =
Configuration error
SUBSTATUS =
Configuration
error
S-
901
Device status
message (FF):
Ambient
temperature too
low
S-901
Local display:
S901 ↔ Measured
value
BLOCK_ERR = Other
QUALITY =
UNCERTAIN
(configurable)
Cause of error:
Reference
temperature <
–40 °C (–40 °F)
Remedy:
Observe ambient
temperature in
accordance with
specification.
SV1, SV2, PV1,
PV2, DT
Transducer_Error = No
error
SUBSTATUS = Non
specific