Operation via FOUNDATION Fieldbus
TM
iTEMP TMT85
92
Hauser
Proceed as follows to change the diagnostic behavior of an event group:
1. Open the assignment parameter in which the group is currently assigned.
2. Change the event group bit from 1 to 0. In the configuration systems, this is done by
deselecting the corresponding check box.
3. Open the assignment parameter to which the group should be assigned.
4. Change the event group bit from 0 to 1. In the configuration systems, this is done by
ticking the corresponding check box.
Example
The Highest Weighting/Configuration Error group contains the events 431: Calibration
values and 437: Configuration error. These should be categorized as Function check (C)
and no longer as Failure (F).
In the Resource Block, search for the "Highest Configuration" group in the FD_FAIL_MAP
parameter and deselect the corresponding check box.
A0042929
Then search for the "Highest Configuration" group in the FD_CHECK_MAP parameter and
tick the corresponding check box.
Summary of Contents for iTEMP TMT85
Page 2: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ...www addresses endress com 71498847 71498847 ...