Proline 400 HART
Electrical connection
Hauser
13
Sensor cable for sensor - transmitter
Standard cable
• TPE: -40 to +80 °C (-40 to +176 °F)
• TPE halogen-free: -40 to +80 °C (-40 to +176 °F)
• PTFE: -40 to +130 °C (-40 to +266 °F)
Cable length (max.)
30 m (90 ft)
Cable lengths (available for order)
5 m (15 ft), 10 m (30 ft), 15 m (45 ft), 30 m (90 ft)
Operating temperature
Depends on the device version and how the cable is installed:
Standard version:
• Cable, fixed installation
: minimum –40 °C (–40 °F)
• Cable, movable: minimum –25 °C (–13 °F)
1)
Compare details under the "Standard cable" row
Cable diameter
• Cable glands supplied:
• For standard cable: M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
• For reinforced cable: M20 × 1.5 with cable 9.5 to 16 mm (0.37 to 0.63 in)
• (Plug-in) spring terminals for wire cross-sections 0.5 to 2.5 mm
2
(20 to 14 AWG)
5.2.3
Terminal assignment
Transmitter
The sensor can be ordered with terminals.
Connection methods available
Possible options for order code
"Electrical connection"
Outputs
Power
supply
Terminals
Terminals
• Option
A
: coupling M20x1 M20x1.5
• Option
B
: thread M20x1 M20x1.5
• Option
C
: thread G ½"
• Option
D
: thread NPT ½"
Supply voltage
Order code
"Power supply"
Terminal numbers
terminal voltage
Frequency range
Option
L
(wide range power unit)
1 (L+/L), 2 (L-/N)
DC 24 V
±25%
–
AC 24 V
±25%
50/60 Hz, ±4 Hz
AC 100 to 240 V
–15 to +10%
50/60 Hz, ±4 Hz
Summary of Contents for HART Proline 400