Conducal CLY421
Troubleshooting
Hauser
19
6.2
Spare parts
501 Device open
M
Close the housing and tighten the screws.
504 Added new user
C
Message relating to changes in the user administration
505 User deleted
C
506 Changed user data
C
510 Invalid parameters
F
Check your settings and adjust them.
513 InternCFW (xxxxxxxx)
F
Contact the Service Team. Quote the error number and the text
displayed.
The (xxxxxxxx) here stands for the text actually displayed.
514 InternCFW (xxxxxxxx)
M
531 (Logbook): full
M
The ring memory of the specified logbook is full. From now on,
new events will overwrite the oldest entries.
810 PV upper limit
F
– Sensor in air
– Air pockets in assembly
– Check measuring chain
PV = Primary value
811 PV lower limit
F
812 Temp upper limit
F
840 PV upper limit
M
– Sensor in air
– Air pockets in assembly
– Check measuring chain
841 PV lower limit
M
842 Temp upper limit
M
843 Temp lower limit
M
950 Conc temp lower limit
M
– Check your process conditions.
– Adjust the measuring range where necessary.
These messages only apply for the concentration tables
saved at the factory.
These messages do not appear if you use user-defined tables.
951 Conc temp upper limit
M
952 Conc kappa lower limit
M
953 Conc kappa upper limit
M
954 Conc lower limit
M
955 Conc upper limit
M
956 Cond temp lower limit
M
957 Cond temp upper limit
M
958 Cond kappa lower limit
M
959 Cond kappa upper limit
M
960 Cond kappa comp lower limit
M
961 Cond kappa comp upper limit
M
Name and content
Order number
for
spare parts kit
Flow assembly complete with carrier, EPDM seals
71148156
Hose connection adapter, locking clamp, EPDM clamp seal
71148157
G1 adapter / clamp
71148158
3 x dummy plugs + 10 x O-rings (EPDM)
71148159
3 x EPDM clamp seals DN20
71148160
4 x Bulgin protection caps (for connecting measuring cables and power supply)
71148161
No. Message text
Cat.
Tests and/or remedial measures