Installation
Condumax CLS15D/16D/21D
14
Hauser
2
1
3
35
.38
(1
)
A0019015
1
With NPT ½" thread in T-piece or cross
fitting
1
T-piece or cross fitting (DN 32, 40 or 50)
2
Glue-in VC threaded coupling (NPT ½" for
DN 20)
3
Glue-in adapter coupling (for DN 32, 40, 50)
NPT ½”
160 (6.30)
Ø 29 (1.14)
Ø 6 (0.24)
30 (1.18)
25 (0.98)
Ø 6 (0.24)
38.5
(1.52)
1
2
3
A0019014
2
With NPT ½" thread in flow assembly
71042405, dimensions in mm (inch)
1
Sensor holder NPT ½"
2
Inlet
3
Outlet
1. Ensure that the electrodes are fully immersed in the medium during measurement.
Immersion depth: at least 35 mm (1.38").
2. If using the sensor in the ultrapure water range, you must work under air-evacuated
conditions.
Otherwise, the CO
2
in the air can dissolve in the water and its (weak) dissociation
can increase the conductivity by up to 3 µS/cm.