3.1.3
Immersion depths
a0017745
Fig. 3:
Immersion depths for different service chambers
1
Single chamber / 36 mm stroke / sensor 225 mm incl. KCl
2
Single chamber / 78 mm stroke / sensor 225 mm excl. KCl
3
Single chamber / 78 mm stroke / sensor 360 mm incl. KCl
4
Double chamber / 78 mm stroke / sensor 225 mm incl. KCl / service position, inner chamber
5
Double chamber / 78 mm stroke / sensor 360 mm excl. KCl / service position, inner chamber
6
Double chamber / 78 mm stroke / sensor 360 mm excl. KCl / service position, front chamber
Immersion depths in mm (inch)
Service chamber
Process connection
1
2
3
4
5
6
CA
Clamp ISO2852
ASME BPE-2012 1½"
X1
X2
20.6 (0.81)
31.6 (1.24)
62.1 (2.44)
73.1 (2.88)
62.1 (2.44)
73.1 (2.88)
28.1 (1.11)
73.1 (2.88)
28.1 (1.11)
73.1 (2.88)
62.1 (2.44)
73.1 (2.88)
CB
Clamp ISO2852
ASME BPE-2012 2"
X1
X2
16.1 (0.63)
27.1 (1.07)
57.6 (2.27)
68.6 (2.70)
57.6 (2.27)
68.6 (2.70)
23.6 (0.93)
68.6 (2.70)
23.6 (0.93)
68.6 (2.70)
57.6 (2.27)
68.6 (2.70)
CC
Clamp ISO2852
ASME BPE-2012 2½"
X1
X2
22.6 (0.89)
33.6 (1.32)
64.1 (2.52)
75.1 (2.96)
64.1 (2.52)
75.1 (2.96)
30.1 (1.19)
75.1 (2.96)
30.1 (1.19)
75.1 (2.96)
64.1 (2.52)
75.1 (2.96)
DA
Aseptic DN 25
Clampable DIN11864-3
A
X1
X2
19.6 (0.77)
30.6 (1.20)
61.1 (2.41)
72.1 (2.84)
61.1 (2.41)
72.1 (2.84)
27.1 (1.07)
72.1 (2.84)
27.1 (1.07)
72.1 (2.84)
61.1 (2.41)
72.1 (2.84)
DC
Aseptic DN 50
Screw-in DIN11864-1 A
X1
X2
27.1 (1.07)
38.1 (1.50)
68.6 (2.70)
79.6 (3.13)
68.6 (2.70)
79.6 (3.13)
34.6 (1.36)
79.6 (3.13)
34.6 (1.36)
79.6 (3.13)
68.6 (2.70)
79.6 (3.13)
DF
Aseptic DN 50
Grooved flange DIN
11864-2 A
X1
X2
21.4 (0.84)
32.4 (1.28)
62.9 (2.48)
73.9 (2.91)
62.9 (2.48)
73.9 (2.91)
28.9 (1.14)
73.9 (2.91)
28.9 (1.14)
73.9 (2.91)
62.9 (2.48)
73.9 (2.91)
EA
Neumo Biocontrol
D 65
X1
X2
27.6 (1.09)
38.6 (1.52)
69.1 (2.72)
80.1 (3.15)
69.1 (2.72)
80.1 (3.15)
35.1 (1.38)
80.1 (3.15)
35.1 (1.38)
80.1 (3.15)
69.1 (2.72)
80.1 (3.15)
EB
Neumo Bioconnect
D 50
X1
X2
22.6 (0.89)
33.6 (1.32)
64.1 (2.52)
75.1 (2.96)
64.1 (2.52)
75.1 (2.96)
30.1 (1.19)
75.1 (2.96)
30.1 (1.19)
75.1 (2.96)
64.1 (2.52)
75.1 (2.96)
EF
Neumo Bioconnect
D 65
X1
X2
20.6 (0.81)
31.6 (1.24)
62.1 (2.44)
73.1 (2.88)
62.1 (2.44)
73.1 (2.88)
28.1 (1.11)
73.1 (2.88)
28.1 (1.11)
73.1 (2.88)
62.1 (2.44)
73.1 (2.88)
MA
Dairy coupling
DN 50 DIN11851
X1
X2
21.1 (0.83)
32.1 (1.26)
62.6 (2.46)
73.6 (2.90)
62.6 (2.46)
73.6 (2.90)
28.6 (1.13)
73.6 (2.90)
28.6 (1.13)
73.6 (2.90)
62.6 (2.46)
73.6 (2.90)
MB
Dairy coupling
DN 65 DIN11851
X1
X2
21.8 (0.86)
32.8 (1.29)
63.3 (2.49)
74.3 (2.93)
63.3 (2.49)
74.3 (2.93)
29.3 (1.16)
74.3 (2.93)
29.3 (1.16)
74.3 (2.93)
63.3 (2.49)
74.3 (2.93)
NA
Thread ISO228
G 1¼
X1
X2
70.4 (2.77)
81.4 (3.20)
70.4 (2.77)
81.4 (3.20)
VA
Varivent flange
N (DN 40 to DN 100)
X1
X2
16.6 (0.65)
27.6 (1.09)
58.1 (2.29)
69.1 (2.72)
58.1 (2.29)
69.1 (2.72)
24.1 (0.95)
69.1 (2.72)
24.1 (0.95)
69.1 (2.72)
58.1 (2.29)
69.1 (2.72)