Commissioning
ASP station 2000 peristaltic
114
Hauser
Thermostat parameters
The parameters for regulating the temperature in the sample compartment are specified in the
"Thermostat" section.
The thermostat is switched on and off by means of the "Operation" box.
The "Target temp." slide control is used to set the desired sample compartment temperature
The defrost time can be set with the "Defrost" slide control. Defrosting is started at 4 a.m. every day
and is active for the set time. If the refrigeration system is active at least 6 hours by 4 p.m., a defrost
cycle is started again at 4 p.m.
Fig. 61:
Transmit/read thermostat set-up (right in screenshot)
The thermostat settings are sent to the controller by means of the "Transmit set-up" button.
Provided that a connection to the controller could be established when the screen was opened, the
settings are read out once at this time. To read out the settings at any time, the "Read set-up" button
can be activated.
Service button
The Service button is protected by a service code and is reserved for use by E+H Service!