For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (M) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
NOTE 3: For a rated maximum output power of transmitter below 20W use a minimum separation distance of
1.5meters for frequencies from 150kHz to 800MHz.
Reader Frequency Band
The myCordella Reader receives RF electromagnetic energy and
includes RF transmitters that perform within the frequency range of
12.8MHz to 14.13MHz.
The myCordella Reader includes RF transmitters that perform at
center bands (13.09MHz, 13.34MHz, 13.62MHz, and 13.9MHz).
BT Transceiver frequency
BT Smart Ready Module
–
FCC ID QOQBT121
Operating Frequency Range: 2402MHz
–
2480MHz
FCC Statement
The equipment is approved for wireless transmission under FCC ID
2AR87ETXCPAS01. It has been tested and found to comply with the
limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates, uses and can radiate radio frequency energy and, if not
used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
•
Increase the separation between the equipment and
receiver.
•
Connect the equipment into an outlet on a circuit different
from that to which the receiver is connected.
•
Consult customer service.