ENDEVCO
www.endevco.com Tel: +1 (866) ENDEVCO [+1 (866) 363-3826]
IM133, Page 16
®
This special mode is selected by pressing the “Select Function” push button while simultaneously
pressing the LEFT EDIT push button. The unit returns to the normal mode of operation (Monitoring
or Non-monitoring state) after pressing the “Select Function” push buttons one more time.
After the mode is selected, all of the “Function LEDs” will be off and the first error code will be
shown on the LED display. The user may scroll through the list by using the UP and DOWN EDIT
push buttons.
3. Reset
This special mode selection allows the user to reset the unit (equivalent to a power-up initialization).
This special mode is selected by pressing the “Select Function” push button while simultaneously
pressing the UP EDIT push button. The unit returns to the normal mode of operation (Monitoring
or Non-monitoring state) immediately after initialization.
4. Store Setup in EEPROM
This special mode selection allows the user to force the unit to store all of the function settings into
non-volatile memory.
This special mode is selected by pressing the “Select Function” push button while simultaneously
pressing the DOWN EDIT push button. The unit returns to the normal mode of operation
(Monitoring or Non-monitoring state) immediately after writing all of the function settings into non-
volatile memory.
5. Test Display
This special mode selection allows the user to test the display board by displaying a different
pattern on the LED indicators and the LED display every time any one of the front panel push
buttons is pressed.
This special mode is selected by pressing the “Select Function” push button while simultaneously
pressing the UP and LEFT EDIT push buttons. The unit returns to the normal mode of operation
(Monitoring or Non-monitoring state) after pressing the “Select Function” push button one more
time.
6. Recall Default Setup
This special mode selection allows the user to recall the default function settings.
This special mode is selected by pressing the “Select Function” push button while simultaneously
pressing the DOWN and RIGHT EDIT push buttons. The unit returns to the normal mode of
operation (Monitoring or Non-monitoring state) immediately after restoring all of the default function
settings.
Function
Factory Default Setting
1. Input Select
Volt (ISOTRON)
2. Current Exc. (mA)
0.0 (off)
3. Sensitivity
1.000
4. Output Scaling
1.000
5. HP Filter
10.0 (on)
6. LP Filter
XX.XX (on)
7. Monitoring State
V out