IM133, Page 15
®
2.4.5
Special Modes
The special modes are not frequently used and they may be selected by pressing the “Select
Function” push button while simultaneously pressing a combination of multiple EDIT push buttons.
The unit must be in the normal mode of operation to enter any of the special modes of operation.
The table below shows the combination of the EDIT push buttons that need to be pressed to obtain
the desired special mode.
Mode
EDIT push-buttons
1.
Addressing
RIGHT
2.
Error List
LEFT
3.
Reset
UP
4.
Store Setup in EEPROM
DOWN
5.
Test Display
UP & LEFT
6.
Recall Default Setup
DOWN & RIGHT
7.
Calibration
LEFT & RIGHT
8.
Troubleshooting
UP & DOWN
9.
Not Assigned
UP & RIGHT
10. Not Assigned
DOWN & LEFT
1. Addressing
This special mode selection provides for programming the address of the unit used for computer
control through the RS-232 serial communication. The unit’s address applies to all 3 channels
(same address regardless of the channel selection).
The unit determines its address at power up by using the following algorithm:
1)
VID: Address determined by ID voltage set in hardware
2)
MEM: Address stored in non-volatile memory
3)
READ VID
4)
IF VID = 0 THEN Address = MEM
5)
IF VID > 0 THEN Address = VID = MEM
This special mode is selected by pressing the “Select Function” push button while simultaneously
pressing the RIGHT EDIT push button. The unit returns to the normal mode of operation
(Monitoring or Non-monitoring state) after pressing the “Select Function” push button one more
time.
When this special mode is selected the first "Function LED" will be blinking and the serial
communication baud rate for the unit will be shown on the LED display. The baud rate can be
modified by using the EDIT buttons. An additional Press of the "Select Function" button will make
the second "Function LED" blink and the address of the unit will be shown on the LED display.
The address can be modified by using the EDIT buttons only if VID = 0. The new address will be
set and stored into the internal non-volatile memory when the unit returns to the normal mode of
operation.
2. Error List
This special mode selection allows the user to retrieve a listing of all of the errors that may be
causing the failure LED to be ON.