17

CLICK!

CLICK!

连接至牙科装置 

  涡轮连接和连接器必须绝对干燥。 连接处潮湿可能导致装置的空气/牙粉通道堵塞。

FB-543_CN_rev_C_Handy 3 Perio.indd   17

03/12/2018   13:57:36

Summary of Contents for AIR-FLOW handy 3.0 PERIO

Page 1: ...AIR FLOW handy 3 0 PERIO Premium AIR FLOW handy 3 0 PERIO AIR FLOW device for dental prophylaxis 20162551590 OPERATION INSTRUCTIONS FB 543_CN_rev_C_Handy 3 Perio indd 1 03 12 2018 13 57 26...

Page 2: ...1a 3 4 5 8 9 10 1b 11 12 6 7 13 2 FB 543_CN_rev_C_Handy 3 Perio indd 2 03 12 2018 13 57 27...

Page 3: ...S 2 Body 3 Powder chamber 4 Powder chamber cap 5 Cord adapter 6 Nozzle remover 7 Nozzles 8 Cap ring 9 Air tube 10 Air powder tube 11 O rings for handpiece connection 12 Nozzle 13 EasyClean EMS supplie...

Page 4: ...configuration with the AIR FLOW PERIO handpiece allows the device to remove subgingival dental plaque from periodontal pockets deeper than 5mm up to 10mm by biofilm removal The configuration with the...

Page 5: ...EMS AIR FLOW handy 3 0 PERIO AIR FLOW handy 3 0 PERIO Premium 2 AIR FLOW PERIO 5mm 10mm AIR FLOW PLUS 4mm FB 543_CN_rev_C_Handy 3 Perio indd 5 03 12 2018 13 57 27...

Page 6: ...arefully as they explain all of the most important details and procedures Please pay special attention to the safety precautions Always keep this instruction close to hand To prevent injury to people...

Page 7: ...FB 543_CN_rev_C_Handy 3 Perio indd 7 03 12 2018 13 57 28...

Page 8: ...ems Please call directly your approved EMS repair center or your dealer We wish you lots of success EMS Technical modifications EMS reserves the right to modify the technique accessories operating ins...

Page 9: ...EMS EMS EMS FB 543_CN_rev_C_Handy 3 Perio indd 9 03 12 2018 13 57 28...

Page 10: ...he dental unit 16 Setting the water flow rate 18 Filling the powder chamber with EASY FILL 20 Closing the cap 22 Precautions 24 Wear a mask and eye protection 24 AIR FLOW PERIO treatment 28 Power sett...

Page 11: ...FLOW PERIO 3 3 3 AIR FLOW PLUS 5 5 5 9 9 9 11 13 13 15 15 17 19 EASY FILL 21 23 25 25 AIR FLOW PERIO 29 29 33 39 AIR FLOW PLUS 41 43 47 49 51 53 55 FB 543_CN_rev_C_Handy 3 Perio indd 11 03 12 2018 13...

Page 12: ...g of the device 62 Daily cleaning of the cap 64 Maintenance 66 Replacing the cap and the O rings 66 Annual maintenance 68 Safety precautions 70 Storing the product 72 Product disposal 74 Warranty 74 A...

Page 13: ...57 57 59 61 63 65 67 O 67 69 71 73 75 75 75 EMS 77 79 83 86 FB 543_CN_rev_C_Handy 3 Perio indd 13 03 12 2018 13 57 28...

Page 14: ...y neutropenia agranulocytosis diabetes haemophilia patients under treatment radiotherapy chemotherapy antibiotics Do not useAIR FLOW PERIO handpiece with nozzle immediately after subgingival cleaning...

Page 15: ...3 AIR FLOW PERIO AIR FLOW PERIO AIR FLOW handy 3 0 PERIO AIR FLOW handy 3 0 PERIO Premium 5mm 10mm AIR FLOW PERIO EMS EMS EMS FB 543_CN_rev_C_Handy 3 Perio indd 3 03 12 2018 13 57 30...

Page 16: ...use of supragingival powders EMS CLASSIC CLASSIC COMFORT and SOFT is strictly forbidden during subgingival treatments contacts with soft tissues dentine cementum Patients suffering from chronic bronc...

Page 17: ...5 AIR FLOW PLUS AIR FLOW PLUS AIR FLOW handy 3 0 PERIO AIR FLOW handy 3 0 PERIO Premium AIR FLOW PLUS 4mm EMS CLASSIC CLASSIC COMFORT SOFT FB 543_CN_rev_C_Handy 3 Perio indd 5 03 12 2018 13 57 30...

Page 18: ...powder i e silica or alumina based This would damage the unit The powders provided by EMS are especially designed for use with the unit Do not use powders from other manufacturers as this could damage...

Page 19: ...7 AIR FLOW EMS EMS FB 543_CN_rev_C_Handy 3 Perio indd 7 03 12 2018 13 57 31...

Page 20: ...supply Water Pressure 1 0 bar P 2 2 bar 1000 2200 hPa Water temperature max 40 C Compressed air supply Air pressure 2 7 P 3 5 bar 2700 3500 hPa Use only dry and clean air oil free FB 543_CN_rev_C_Hand...

Page 21: ...9 1 0 P 2 2 1000 2200 hPa 40 C 2 7 P 3 5 2700 3500 hPa FB 543_CN_rev_C_Handy 3 Perio indd 9 03 12 2018 13 57 32...

Page 22: ...to another type of turbine will damage it The turbine of your dental unit must not be under pressure when to connect the unit Do not activate the turbine footpedal If you have a turbine equipped with...

Page 23: ...11 O O FB 543_CN_rev_C_Handy 3 Perio indd 11 03 12 2018 13 57 32...

Page 24: ...maged The single use nozzle must be used for one single patient only Never reuse a nozzle because treatment will be ineffective and the risk of emphysema will increase The nozzle cannot be sterilized...

Page 25: ...13 FB 543_CN_rev_C_Handy 3 Perio indd 13 03 12 2018 13 57 34...

Page 26: ...ore first use and after each treatment as described in the chapter Cleaning disinfecting and sterilizing in this manual Following the thermal disinfection blow the interior of the handpiece from both...

Page 27: ...15 FB 543_CN_rev_C_Handy 3 Perio indd 15 03 12 2018 13 57 35...

Page 28: ...dental unit The turbine connection and the connector must be absolutely dry Moisture at the connection can cause blockage in the air powder channels of the unit FB 543_CN_rev_C_Handy 3 Perio indd 16...

Page 29: ...17 CLICK CLICK FB 543_CN_rev_C_Handy 3 Perio indd 17 03 12 2018 13 57 36...

Page 30: ...water flow It is better to set the water flow rate before using the device for the first time while the powder chamber is empty Set the water flow rate from dental unit to obtain a small continuous s...

Page 31: ...19 FB 543_CN_rev_C_Handy 3 Perio indd 19 03 12 2018 13 57 36...

Page 32: ...t the powder chamber and the cap of the powder chamber are absolutely dry Moisture can cause the powder to cake Use only powder provided by EMS Do not go over the maximum limit as indicated on the pic...

Page 33: ...21 1 2 MAX EASY FILL EMS FB 543_CN_rev_C_Handy 3 Perio indd 21 03 12 2018 13 57 38...

Page 34: ...efore screwing on the cap Ensure that the cap is correctly screwed to avoid projection of powder Do not shake the device as this could cause the powder to clog the tubes Check that you get a water spr...

Page 35: ...23 FB 543_CN_rev_C_Handy 3 Perio indd 23 03 12 2018 13 57 40...

Page 36: ...nhalation of powder we recommend to the dentist and the hygienist to wear gloves protective glasses and a safety mask of protection The contact lenses or glasses of the patient may get dirty during th...

Page 37: ...25 FB 543_CN_rev_C_Handy 3 Perio indd 25 03 12 2018 13 57 41...

Page 38: ...y prevent bacteria development during the treatment BacterX pro is not available in the USA Application of Vaseline on patient s lips will prevent them becoming dry and chapped The cotton rolls stop t...

Page 39: ...27 m a c h o ire 3 BacterX pro 30 BacterX pro FB 543_CN_rev_C_Handy 3 Perio indd 27 03 12 2018 13 57 42...

Page 40: ...t at medium setting Then adjust the irrigation to obtain a flow as required for treatment Start the treatment with the power setting air powder control to the minimum setting It can then be adjusted a...

Page 41: ...29 AIR FLOW PERIO PERIO FLOW FB 543_CN_rev_C_Handy 3 Perio indd 29 03 12 2018 13 57 42...

Page 42: ...o ire 3 The handpiece and the high speed evacuator must be handled by the same operator The high speed evacuator is thus directed in an optimal way towards the nozzle FB 543_CN_rev_C_Handy 3 Perio in...

Page 43: ...31 FB 543_CN_rev_C_Handy 3 Perio indd 31 03 12 2018 13 57 43...

Page 44: ...kets down to the apex At least 3 mm of bone must be present to prevent the risk of emphysema Verify with radiographs pre treatment Do not use the PERIO FLOW method immediately after a sub gingival sca...

Page 45: ...33 5mm AIR FLOW handy 3 0 PERIO 3mm X PERIO FLOW FB 543_CN_rev_C_Handy 3 Perio indd 33 03 12 2018 13 57 44...

Page 46: ...ds 3 Place the nozzle inside the pocket for 5 seconds maximum per site Repeated insertion of the nozzle in the same site is not recommended Carry on the treatment with short tiny vertical motions of t...

Page 47: ...35 3 1 2 5 3 5 EMS FB 543_CN_rev_C_Handy 3 Perio indd 35 03 12 2018 13 57 44...

Page 48: ...more seconds Take these seconds into account 5 Remove the nozzle from the pocket and repeat the operation in the next pocket Never spray outside periodontal pockets External operation accelerates nozz...

Page 49: ...37 5 4 5 FB 543_CN_rev_C_Handy 3 Perio indd 37 03 12 2018 13 57 45...

Page 50: ...e by rotating the handpiece while pulling the handpiece along the axis Remove the nozzle as shown To remove the nozzle with the nozzle remover please refer to the PERIO FLOW Nozzle Remover user manual...

Page 51: ...39 135 C PERIO FLOW FB 543_CN_rev_C_Handy 3 Perio indd 39 03 12 2018 13 57 47...

Page 52: ...ental unit to evacuate the air powder mixture deviated by the treated tooth The handpiece and the high speed evacuator must be handled by the same operator The high speed evacuator is thus directed in...

Page 53: ...41 AIR FLOW PLUS FB 543_CN_rev_C_Handy 3 Perio indd 41 03 12 2018 13 57 48...

Page 54: ...en nozzle and tooth from 30 to 60 degrees The flatter the angle the larger the area cleaned During treatment direct the high speed pump towards the air powder jet deviated by the tooth The angle of re...

Page 55: ...43 3 5 mm 30 60 FB 543_CN_rev_C_Handy 3 Perio indd 43 03 12 2018 13 57 48...

Page 56: ...tissue spaces It is strongly recommended that the operator never direct the nozzle directly at the gum tissue or into the gingival sulcus During treatment make small circular motions Finally after th...

Page 57: ...45 FB 543_CN_rev_C_Handy 3 Perio indd 45 03 12 2018 13 57 49...

Page 58: ...of the tooth Respect a distance of 3 to 5 mm During treatment make small horizontal or circular motions You can vary the angle between 30 to 60 degrees The flatter the angle the larger the area clean...

Page 59: ...47 3 5 mm 30 60 5 FB 543_CN_rev_C_Handy 3 Perio indd 47 03 12 2018 13 57 50...

Page 60: ...jet continues for a few more seconds End the treatment taking this time frame into account The nozzle can be introduced into the high speed evacuator still in the patient mouth This allows time for th...

Page 61: ...49 FB 543_CN_rev_C_Handy 3 Perio indd 49 03 12 2018 13 57 50...

Page 62: ...tient can do a final rinse After the treatment the teeth are practically free from mucin It is thus advised to carry out a topics application of fluorine It is then important to use a colorless fluori...

Page 63: ...51 FB 543_CN_rev_C_Handy 3 Perio indd 51 03 12 2018 13 57 51...

Page 64: ...ins requires between 2 and 3 hours During this time the teeth do not have any more natural protection with respect to colouring Tell your patient that during 2 to 3 hours following treatment he should...

Page 65: ...53 2 3 2 3 FB 543_CN_rev_C_Handy 3 Perio indd 53 03 12 2018 13 57 52...

Page 66: ...zations must be carried out immediately no later than 30 minutes after the end of the treatment To easily disconnect the handpiece from the body rotate slightly the handpiece while pulling in the axis...

Page 67: ...55 30 FB 543_CN_rev_C_Handy 3 Perio indd 55 03 12 2018 13 57 52...

Page 68: ...g the handpiece Clean the handpiece with the EasyClean supplied with the unit before disinfecting and sterilizing CLEANING DISINFECTING AND STERILIZING FB 543_CN_rev_C_Handy 3 Perio indd 56 03 12 2018...

Page 69: ...57 EasyClean FB 543_CN_rev_C_Handy 3 Perio indd 57 03 12 2018 13 57 53...

Page 70: ...SA law restri cts this devic e to sale by or on the order of a denti st CLEANING DISINFECTING AND STERILIZING Do not exceed the maximum number of sterilization cycles St rilization of the handpiece FB...

Page 71: ...59 FB 543_CN_rev_C_Handy 3 Perio indd 59 03 12 2018 13 57 53...

Page 72: ...y available ethanol isopropanol colorless disinfectant The use of scouring powder or an abrasive sponge will damage its surface Do not submerge the unit in a disinfectant bath It would be likely to be...

Page 73: ...61 135 C FB 543_CN_rev_C_Handy 3 Perio indd 61 03 12 2018 13 57 55...

Page 74: ...ork day the powder chamber Empty the powder chamber Suck the powder residues with the high speed evacuator of the dental unit Clean the threads of the powder chamber with alcohol ethanol isopropanol F...

Page 75: ...63 FB 543_CN_rev_C_Handy 3 Perio indd 63 03 12 2018 13 57 56...

Page 76: ...the end of the work day The cap must be absolutely dry before closing To ensure a good tightness the cap ring must be replaced at least once a year Daily cleaning of the cap FB 543_CN_rev_C_Handy 3 P...

Page 77: ...65 FB 543_CN_rev_C_Handy 3 Perio indd 65 03 12 2018 13 57 56...

Page 78: ...powder chamber is pressurized during use The condition of the powder chamber and cap is an important safety factor Replace faulty parts immediately If an o ring is damaged or worn out it must be repl...

Page 79: ...67 O O FB 543_CN_rev_C_Handy 3 Perio indd 67 03 12 2018 13 57 58...

Page 80: ...ce To keep your device in a perfect operating and to extend its life send your device each year for an annual maintenance to EMS or an approved EMS repair center FB 543_CN_rev_C_Handy 3 Perio indd 68...

Page 81: ...69 EMS EMS FB 543_CN_rev_C_Handy 3 Perio indd 69 03 12 2018 13 57 59...

Page 82: ...ces and accessories may cause bacterial or viral infections Use only the powder recommended by EMS for the subgingival application when using the AIR FLOW handy PERIO Any other powder could cause prem...

Page 83: ...71 EMS AIR FLOW handy PERIO EMS EMS EMS EMS EMS FB 543_CN_rev_C_Handy 3 Perio indd 71 03 12 2018 13 58 01...

Page 84: ...ny time Should you wish to store the product for an extended period of time Proceed as described in the chapter Cleaning disinfecting and sterilizing Pack the product and all accessories in the origin...

Page 85: ...73 FB 543_CN_rev_C_Handy 3 Perio indd 73 03 12 2018 13 58 01...

Page 86: ...ply in your country WARRANTY Damages due to non adherence to the operating instructions or wear out of parts are excluded from warranty The warranty of your product will be cancelled if you try to ope...

Page 87: ...75 EMS FB 543_CN_rev_C_Handy 3 Perio indd 75 03 12 2018 13 58 02...

Page 88: ...t against damage during shipment Before dispatching your product including all accessories please clean disinfect and sterilize as described in the operating instructions The connector must be dismant...

Page 89: ...77 EMS EMS EMS EMS EMS EMS FB 543_CN_rev_C_Handy 3 Perio indd 77 03 12 2018 13 58 02...

Page 90: ...ilizable at up to 135 C in the autoclave CE marking Refers to directive 93 42 EEC including EN 60601 1 and EN 60601 1 2 Serial number Catalogue number Lot number Manufacturer Manufacture date FB 543_C...

Page 91: ...79 135 C 135 C CE 93 42 EEC EN 60601 1 EN 60601 1 2 FB 543_CN_rev_C_Handy 3 Perio indd 79 03 12 2018 13 58 03...

Page 92: ...80 Thermal disinfection Refer to instruction manual FB 543_CN_rev_C_Handy 3 Perio indd 80 03 12 2018 13 58 03...

Page 93: ...81 FB 543_CN_rev_C_Handy 3 Perio indd 81 03 12 2018 13 58 03...

Page 94: ...the turbine connection Air Static pressure 2 7 to 3 5 bar 2700 3500 hPa Weight Approx 0 150 kg Operating conditions 10 C to 40 C 30 to 75 relative humidity 700 hPa to 1060 hPa air pressure Altitude ma...

Page 95: ...handy 3 0 PERIO Premium 93 42 EEC IIa 1 2 2 1000 2200 hPa 50 80 ml 2 7 3 5 2700 3500 hPa 0 150 kg 10 C 40 C 30 75 700 hPa 1060 hPa 3000 m 800 5 40 C 5 85 500 hPa 1060 hPa 10 C 40 C 60 C 75 700 hPa FB...

Page 96: ...pedal of the dental unit If air comes through the handpiece connector the handpiece is blocked Clean the handpiece If no air comes out of the handpiece connector the unit is blocked Send the unit to...

Page 97: ...r If the handpiece falls AIR FLOW PERIO handpiece Replace the nozzle and check the fit of the nozzle If the nozzle doesn t hold or it is not watertight send the handpiece to a certified EMS Repair Cen...

Page 98: ...86 O EMS 1 2 EMS EMS EMS O EMS EMS FB 543_CN_rev_C_Handy 3 Perio indd 86 03 12 2018 13 58 04...

Page 99: ...87 EasyClean EMS AIR FLOW PERIO EMS AIR FLOW PLUS Easyfill FB 543_CN_rev_C_Handy 3 Perio indd 87 03 12 2018 13 58 04...

Page 100: ...FB 543_CN_rev_C_Handy 3 Perio indd 88 03 12 2018 13 58 04...

Page 101: ...FB 543_CN_rev_C_Handy 3 Perio indd 89 03 12 2018 13 58 04...

Page 102: ...witzerland 41 22 994 4700 41 22 994 4701 ems dental com contact Rte de Champ Colin 2 1260 Nyon Switzerland 108 2 310 379 24 A 021 33632323 Copyright EMS SA FB 543 CN rev C ed 2018 12 e e Swiss Periodo...

Reviews: