Maintenance

To maintain optimum performance, cleaning should 

be performed on a regular basis. Use 99% isopropyl 

alcohol to clean the foot plate and test switch. Other 

cleaners are susceptible to leaving residue on these 

surfaces.
NOTE: Avoid wetting or mopping the foot plate. This 

may yield false test results and cause damage to its 

internal circuitry.

Calibration

Frequency of recalibration should be based on the 

critical nature of those ESD sensitive items handled and 

the risk of failure for the ESD protective equipment and 

materials. In general, EMIT recommends that calibration 

be performed annually.
Use the EMIT 50424 Limit Comparator and EMIT 50784 

5-Pound Electrode to perform periodic verification (once 

every 6-12 months) of the SmartLog Pro®. The Limit 

Comparator can be used to check the test limits of the 

SmartLog Pro® without removing it from the factory 

floor.
See 

TB-6581

 for more information.

NOTE: Calibration is not required for the temperature 

and humidity sensor embedded within the SmartLog 

Pro®.

Figure 22.  EMIT 50424 Limit Comparator

Figure 23.  EMIT 50784 5-Pound Electrode for Limit 

Comparator

Specifications

SmartLog Pro

®

Input Voltage and 

Frequency

(External Adapter)

AC/DC Power Adapter
Power Input: 

100-240VAC, 50/60 Hz
Power Output: 

5VDC, 3.0A
Cable Length: 

5 ft. (1.5 m)

Operating Temperature 70°F to 85°F (21°C to 30°C) 

for 1 gigohm test limit
41°F to 85°F (5°C to 30°C) 

for all other test limits

Environmental 

Requirements

Indoor use only at altitudes 

less than 6500 ft. (2 km)
Maximum relative humidity 

of 80% up to 85°F (30°C) 

decreasing linearly to 50% @ 

85°F (30°C)
Maximum relative humidity of 

50% at 1 gigohm setting

Dimensions

6.75" x 5.00" x 1.75"

(17.1 cm x 12.7 cm x 4.4 cm)

Weight

1.8 lbs (0.8 kg)

Test Accuracy

±20% for 1 gigohm footwear 

test limit
±10% for all other test limits

Test Switch Voltage

5 VDC @ open circuit

Wrist Strap and 

Footwear Test Voltage

30 VDC @ open circuit
Test current is limited by 

resistors and varies on 

the test range setting (100 

kilohms - 1 gigohm)

Temperature Accuracy ±0.5°C 
Humidity Accuracy

±5%

Dual Independent Foot Plate

Dimensions

14.0" x 16.0" x 0.9"

(35.6 cm x 40.1 cm x 2.3 cm)

Weight

7.5 lbs (3.4 kg)

Limited Warranty, Warranty Exclusions, Limit of 

Liability and RMA Request Instructions

See the EMIT Warranty - 

DescoEMIT.com/Warranty.aspx

Summary of Contents for 50780 SmartLog Pro

Page 1: ...mated reports that may be e mailed to track short term and long term corrective actions Test limits and test parameters wrist strap only footwear only wrist strap and footwear etc may be customized wi...

Page 2: ...DescoEMIT com Contents 3 System Overview 4 Items and Accessories 4 Packaging 5 Features and Components 6 Installation 6 Hardware Setup 7 10mm Wrist Cord Adapter 8 Relay Terminal 8 Network Setup 9 Tes...

Page 3: ...site DescoEMIT com System Overview SmartLog Pro SmartLog Pro PC with TEAM5 Software Supervisor Department Manager Department Manager Department Manager Figure 2 Illustration of the SmartLog Pro System...

Page 4: ...787 Replacement Foot Plate Cable 50788 Handheld QR Code Scanner USB Use the SmartLog Pro along with the TEAM5 Software to automate the collection of employee ESD testing This software is offered in tw...

Page 5: ...12 2018 DESCO INDUSTRIES INC Employee Owned EMIT 3651 Walnut Avenue Chino CA 91710 909 664 9980 Website DescoEMIT com Features and Components Figure 3 SmartLog Pro features and components A B C D E F...

Page 6: ...egrates with electronic door locks lights buzzers etc See Relay Terminal on page 8 for more information N Ground Terminal Secure the tinned wire termination of the included ground cord to this termina...

Page 7: ...face of the SmartLog Pro Be sure to clear the clip at the base of the plunger s shaft when inserting into the jack 6 Connect the SmartLog Pro to the bracket Use the included thumbscrew to secure the...

Page 8: ...tLog Pro to a local area network LAN using its Ethernet communication module DYNAMIC IP PROCEDURE 1 Verify that the Ethernet cable is securely connected to your network and SmartLog Pro The LEDs on th...

Page 9: ...te of this address as it will be required to establish communication between the SmartLog Pro and the TEAM5 Server Software Figure 13 Locating the IP address in the Admin Menu Figure 14 Locating the N...

Page 10: ...When used with the EMIT SmartLog TEAM5 allows manufacturers and assemblers to automate the collection of employee ESD testing It has a set of robust employee management functions that allows automated...

Page 11: ...lean with 99 isopropyl alcohol when using the 1 Gigohm high test limit A dirty foot plate could yield a false pass When performing a wrist strap test be sure to completely plug in the wrist cord into...

Page 12: ...umidity sensor embedded within the SmartLog Pro Figure 22 EMIT 50424 Limit Comparator Figure 23 EMIT 50784 5 Pound Electrode for Limit Comparator Specifications SmartLog Pro Input Voltage and Frequenc...