4310 Wireless Position Monitor
Instruction Manual
October 2009
14
Online
1 Overview
2 Configure
3 Service Tools
Configure
1 Guided Setup
2 Manual Setup
3 Alert Setup
4 Calibration
Manual Setup
1 Device Identification
2 Wireless
3 Burst Mode
4 Broadcast Power
5 Power Information
6 Trigger Points
Device Identification
1 HART Tag
2 HART Long Tag
3 Descriptor
4 Message
Wireless
1 Wireless Mode
2 Network ID
2 Join Key to Network
Burst Mode
1 Burst Mode
2 Update Period
3 Change Update Pe...
4 Show Burst Variables
Broadcast Power
1 Power Level
Power Information
1 Power Source
2 Power Module Type
3 Power Module Date
4 Power Module Volt
5 Remaining Life
6 Reset Power Module
Trigger Points
1 Switch Status
2 Open Trigger Point
3 Closed Trigger Point
4 Deadband (5%)
5 Change Trigger Pts
Cycle Count Alert
1 Alert Trigger
2 Accumulated Cycles
3 Reset Cycle Counter
2
2-3-1
2-2-1
2-2
2-2-2
2-2-3
2-2-5
2-2-4
Field Communicator
1 Offline
2 Online
3 Frequency Device
4 Utility
2-2-6
Guided Setup
1 Device Setup
2-1
Alert Setup
1 Cycle Count Alert
2 Open/Closed Alerts
2-3
Calibration
1 Calibration Status
2 Calibration Date
3 Calibrate
Open/Closed Alerts
1 Setup Open/Closed Alerts
2-4
2-3-2
Figure 16. Field Communicator Menu Tree—Configure
Calibrate
Follow the prompts on the Field Communicator to
calibrate the instrument.
Join Device to Network allows you to enter the
Network ID (up to 5
−
digits) and the Join Key (4 sets
of 8 hexadecimal digits). The Join Key is used for
encrypting the data on the wireless network. The
Network ID and the Join Key must be the same for
all devices connected to a gateway.
Note
The Network ID is shipped with a
default value of 1229 and the Join Key
with a default value of all zeros. These
must be changed prior to use.
Configure
Refer to figure 16 for the Configure menu tree.
Guided Setup
D
Device Setup
Follow the prompts to setup the instrument.
Manual Setup
D
Device Identification
HART Tag—enter an up to 8 character HART tag for
the instrument. The HART tag is the easiest way to
distinguish between instruments in a
multi
−
instrument environment. Use the HART tag to
label instruments electronically according to the
requirements of your application. The tag you assign
is automatically displayed when the Field
Communicator establishes contact with the 4310 at
power
−
up.