Step 1- Mounting the 370XA
Recommended Minimum Installation Clearances
1660 mm
(65 in)
50 mm
(2 in)
635 mm
(25 in)
280 mm
(11 in)
665 mm (38 in)
200 mm
(8 in)
460 mm
(18 in)
275.6 mm
(10.849 in)
1 m
(40 in)
200 mm
(8 in)
200 mm
(8 in)
305 mm
(12 in)
705 mm
(28 in)
1000 mm
(40 in)
Attach the 370XA bracket to the pole by sliding the plastic inserts
onto the pole and bolting the two U-bolts through the inserts and
the holes in the bracket. The base of the bracket should be 40 inches
from the ground to make the operator interface easy to use.
1000 mm
(40 in)
9 in
(230 mm)
11 in
(280 mm)
9 in
(230 mm)
11 in
(280 mm)
Secure to the floor with four ½ - inch or
⅜
- inch cement anchors.
The cement foundation should be able to support the weight of the
GC (45 lbs) and its accessories, as well as any other utilities that may
be installed.
If supplying your own pole, it should be four inches in diameter and
at least 55 inches tall.
Mount the bracket on the pole.