20
Reference Manual
00809-0200-4648, Rev BB
July 2016
To input Callendar-Van Dusen constants, perform the following procedure:
1.
From the
HOME
screen, select
2 Configure, 1 Guided Setup, 1 Configure Sensor,
1 Configure Type and Units
and press
Enter.
2.
Select
Cal VanDusen
at the
Select Sensor Type
prompt.
3.
Select the appropriate number of wires at the
Select Sensor Connection
prompt.
4.
Enter the R
o
, Alpha, Delta, and Beta values from the stainless steel tag attached to
the special-order sensor when prompted.
5.
Select desired other options and select
Enter
.
6.
To disable the Transmitter-sensor matching feature from the
HOME
screen select
Configure>Guided Setup>Configure Sensor>Configure Sensor Type and Units
and press
Enter
. Select the appropriate sensor type from the Select Sensor type prompt.
Note
When the transmitter-sensor matching is disabled, the transmitter reverts to factory trim.
Make certain the transmitter engineering units default correctly before placing the
transmitter into service.
2.9
Advanced setup
2.9.1
LCD display
The LCD display configuration command allows customization of the LCD display to suit
application requirements. The LCD display will alternate between the selected items:
Temperature units
Sensor temperature
% of range
Supply voltage
Reference
“LCD display screen messages” on page 43
for images of LCD display screens.
Summary of Contents for Rosemount X-well 648
Page 2: ......
Page 6: ...vi Reference Manual 00809 0200 4648 Rev BB Title Page July 2016 Title Page ...
Page 10: ...4 Reference Manual 00809 0200 4648 Rev BB Contents July 2016 Contents ...
Page 12: ...2 Reference Manual 00809 0200 4648 Rev BB Introduction July 2016 Introduction ...
Page 36: ...26 Reference Manual 00809 0200 4648 Rev BB Configuration July 2016 Configuration ...
Page 52: ...42 Reference Manual 00809 0200 4648 Rev BB Commissioning July 2016 Commissioning ...
Page 66: ...56 Reference Manual 00809 0200 4648 Rev BB Troubleshooting July 2016 Troubleshooting ...
Page 89: ......