Canada Intrinsically Safe (IS)

Certificate:

[CSA]70047656

Standards:

CAN/CSA C22.2 No. 0-10, CAN/CSA C22.2 No. 94-M1991
(R2011), CAN/CSA-60079-0-11, CAN/CSA-60079-11-14, CSA
Std C22.2 No. 60529-05, CAN/CSA-C22.2 No. 61010-1-12

Markings:

Intrinsically Safe for Class I, Division 1, Groups A, B, C, D T4; Ex
ia IIC T4 Ga T4 (–50 °C ≤ Ta ≤ +70 °C) when installed per
Rosemount drawing 00G45-1020; Type 4X; IP66/67;

Special Conditions for Safe Use (X):

1. Do not replace battery when explosive atmosphere is present. Ne pas

remplacer les accumulateurs si une atmosphère explosive peut être
présente.

2. Use only 00G45-9000-0001 batteries. Utiliser uniquement des

accumulateurs 00G45-9000-0001.

3. The surface resistivity of the housing is greater than 1G

. To avoid

electrostatic charge build-up, it must not be rubbed or cleaned with
solvents or a dry cloth. La résistivité de surface du boÎtier est
supérieure à un gigaohm. Pour éviter l’accumulation de charge
électrostatique, ne pas frotter ou nettoyer avec des produits solvants
ou un chiffon sec.

4. Substitution of components may impair intrinsic safety. La

substitution de composants peut compromettre la sécurité
intrinsèque.

Europe

I1

ATEX Intrinsic Safety

Certificate:

Baseefa16ATEX0005X

Standards:

EN 60079-0: 2012 + A11: 2013, EN 60079-11: 2012

Markings:

 II 1 G Ex ia IIC T4 Ga, T4 (–40 °C ≤ Ta ≤ +70 °C) IP66/67;

Special Conditions for Safe Use (X):

1. The plastic enclosure may constitute a potential electrostatic ignition

risk and must not be rubbed or cleaned with a dry cloth.

2. The measured capacitance between the equipment enclosure and

metallic inline sensor module is 4.7pF. This must be considered only
when the SPG is integrated into a system where the process
connection is not grounded.

Quick Start Guide

February 2019

14

Emerson.com/Rosemount

Summary of Contents for Rosemount WPG: Rosemount SPG

Page 1: ...Quick Start Guide 00825 0100 4145 Rev BB February 2019 Rosemount Smart Pressure Gauge ...

Page 2: ...lt current regulations and requirements before shipping WARNING Explosions could result in death or serious injury Installation of device in an explosive environment must be in accordance with appropriate local national and international standards codes and practices Ensure device is installed in accordance with intrinsically safe or non incendive field practices Electrical shock could cause death...

Page 3: ...1 Required equipment Figure 1 1 Anti seize paste or PTFE tape for NPT threaded connection Figure 1 2 Standard tools e g screwdriver wrench pliers February 2019 Quick Start Guide Quick Start Guide 3 ...

Page 4: ...k Start Guide The following options are also available and will ship with the Rosemount Smart Pressure Gauge if ordered Rosemount 306 Integral Manifold model code S5 B4 bracket model code B4 Quick Start Guide February 2019 4 Emerson com Rosemount ...

Page 5: ...ation Cert and Tamper Evident Seal Q8 Material Traceability cert per EN 102043 1 Q15 Cert of Compliance to NACE MR0175 ISO 15156 for wetted materials Q25 Cert of Compliance to NACE MR00103 for wetted materials Normal range Indication model code LK Product Certification model codes February 2019 Quick Start Guide Quick Start Guide 5 ...

Page 6: ...e is designed to be installation ready Procedure 1 To check device battery prior to installation perform Turn on device 2 Slide the ON OFF switch to the OFF position until ready for use Quick Start Guide February 2019 6 Emerson com Rosemount ...

Page 7: ...ion until it is pointing at the red X NOTICE Equipment Damage Use caution as the electronics assembly is connected to the needle 5 Remove any debris on the dial so it does not become trapped under the sticker 6 Peel back the white paper backing of the sticker 7 Slowly lower the sticker onto the surface of the dial in the desired location and rub it in place firmly Repeat steps Step 6 and Step 7 un...

Page 8: ...e pressure port atmospheric reference is located in the neck of the device behind the housing The vent path is between the housing and sensor CAUTION Keep the vent path free of any obstruction including but not limited to paint dust and lubrication by mounting the device so the process can drain away Quick Start Guide February 2019 8 Emerson com Rosemount ...

Page 9: ...he cover counterclockwise to remove it 2 Slide the OFF ON switch to the ON position to initiate the power sequence Note During the power sequence the dial tests full range of motion and LED flashes amber 3 Once the power sequence ends verify that the LED flashes green The LED may display several colors see Device status indicator February 2019 Quick Start Guide Quick Start Guide 9 ...

Page 10: ...ect Device Connection View 3 Double click the device under the HART modem 5 3 2 Field communicator Procedure to connect the smart pressure gauge to the field communicator Procedure 1 Turn on the Field Communicator 2 From the Main menu select the HART symbol 5 4 Eliminate mounting effects Devices are factory calibrated Once installed it is recommended to perform this step to eliminate potential err...

Page 11: ...gurator Sensor trim function on absolute gauge Procedure 1 Vent the device 2 Connect the Field Communicator 3 From the HOME screen enter the HART Fast Key sequence Device dashboard Fast Keys 2 1 1 4 Follow the commands to perform the procedure February 2019 Quick Start Guide Quick Start Guide 11 ...

Page 12: ...tioning No color No power verify ON OFF switch is in the on position 6 2 Pressure measurement If the mounting effects have not been eliminated after completing perform this alternative procedure for verifying the pressure value Procedure 1 From the HOME screen enter the HART Fast Key sequence Device dashboard Fast Keys 2 2 1 1 1 2 Follow the commands to perform the procedure Quick Start Guide Febr...

Page 13: ...ent in Divisions The markings must be suitable for the area classification gas and temperature class This information is clearly defined in the respective codes USA I5 U S A Intrinsically Safe IS Certificate CSA 70047656 Standards FM 3600 2011 FM 3610 2010 UL Standard 50 Eleventh Edition UL 61010 1 3rd Edition ANSI ISA 60079 0 12 00 01 2013 ANSI ISA 60079 11 12 02 01 2013 ANSI IEC 60529 2004 Marki...

Page 14: ...bbed or cleaned with solvents or a dry cloth La résistivité de surface du boÎtier est supérieure à un gigaohm Pour éviter l accumulation de charge électrostatique ne pas frotter ou nettoyer avec des produits solvants ou un chiffon sec 4 Substitution of components may impair intrinsic safety La substitution de composants peut compromettre la sécurité intrinsèque Europe I1 ATEX Intrinsic Safety Cert...

Page 15: ... 1 The plastic enclosure may constitute a potential electrostatic ignition risk and must not be rubbed or cleaned with a dry cloth 2 The measured capacitance between the equipment enclosure and metallic inline sensor module is 4 7pF This must be considered only when the SPG is integrated into a system where the process connection is not grounded 3 Do not change the battery when an explosive atmosp...

Page 16: ... 9685 USA to which this declaration relates is in conformity with the provisions of the European Community Directives including the latest amendments as shown in the attached schedule Assumption of conformity is based on the application of the harmonized standards and when applicable or required a European Community notified body certification as shown in the attached schedule signature Vice Presi...

Page 17: ...Pressure Gauge only Harmonized Standards EN 300 328 V2 1 1 EN 301 489 1 V2 2 0 EN 301 489 17 V3 2 0 EN 61010 1 2010 EN 62479 2010 ATEX Directive 2014 34 EU Models WPG SPG Baseefa16ATEX0005X Intrinsic Safety Certificate Equipment Group II Category 1 G Ex ia IIC T4 Ga T4 40 C Ta 70 C Harmonized Standards EN 60079 0 2012 A11 2013 EN 60079 11 2012 February 2019 Quick Start Guide Quick Start Guide 17 ...

Page 18: ...s SGS FIMCO OY Notified Body Number 0598 P O Box 30 Särkiniementie 3 00211 HELSINKI Finland ATEX Notified Body for Quality Assurance SGS FIMCO OY Notified Body Number 0598 P O Box 30 Särkiniementie 3 00211 HELSINKI Finland Quick Start Guide February 2019 18 Emerson com Rosemount ...

Page 19: ...bly X O O O O O 㓴Ԧ Battery Assembly X O O O O O ᵜ㺘Ṭ ᦞ SJ T11364Ⲵ㿴ᇊ㘼ࡦ This table is proposed in accordance with the provision of SJ T11364 O Ѫ䈕䜘ԦⲴᡰᴹ 䍘ᶀᯉѝ䈕ᴹᇣ 䍘Ⲵਜ਼䟿 վҾGB T 26572ᡰ㿴ᇊⲴ䲀䟿㾱 O Indicate that said hazardous substance in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Ѫ 䈕䜘Ԧᡰ Ⲵᡰᴹ 䍘ᶀᯉ䟼ˈ㠣ቁᴹа 䍘ᶀᯉѝ䈕ᴹᇣ 䍘Ⲵਜ਼䟿儈ҾGB T 26572ᡰ㿴ᇊⲴ䲀䟿㾱 X Indicate that said hazardo...

Page 20: ... Baar Switzerland 41 0 41 768 6111 41 0 41 768 6300 RFQ RMD RCC Emerson com Asia Pacific Regional Office Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 65 6777 8211 65 6777 0947 Enquiries AP Emerson com Middle East and Africa Regional Office Emerson Automation Solutions Emerson FZE P O Box 17033 Jebel Ali Free Zone South 2 Dubai United Arab Emirates 971 4 8118100 971 4 8865465 RFQ...

Reviews: