c) Make sure that the below parameters are set with the
following values:
•
Propagation Factor: 1,000
•
Tank Connection Length: 0
•
Remote housing: None
•
Probe angle: 0
•
Vapor DC: 1
•
Measurement mode: Liquid product level, solid product
level or product level and interface level
d) Save these configuration changes and restart the device.
2. Perform the verification test.
a) Set the cable switch to position 1. Wait for the values to
stabilize before taking a reading from the
Device Display
window in Rosemount Radar Master.
b) Record the distance value, the internal temperature of
transmitter, and the Rosemount VeriCase temperature.
for all six cable lengths in the
Rosemount VeriCase.