6
5081C-HT
Instruction Sheet
January 2015
LIQ_MAN_ABR_5081C-HT/Rev.G
VARIAble NAMe
MNeMoNIC
FACToRy SeTTINGS
CUSToMeR
SeTTINGS
program Menu
output
Output
–
___________
4 mA
4 mA
0 µS
___________
20 mA
20 mA
20 mS
___________
Hold
hold
21 mA
___________
Fault
fault
22 mA
___________
Dampening
dpn
0 samples/second
___________
Test
test
04.00 mA
___________
Temperature
temp
Auto temperature compensation
tauto
on
___________
Manual temperature
tman
25.0°C (overridden by auto)
___________
Temperature compensation
COMP
(
Linear
or
nOnE
)
LInEAr
___________
algorithm
display
display
Measurement type
typ
(
CondUC
or
RSTVTY
or
Cust
)
Temperature (°C or °F)
temp
C
___________
Output (mA or %)
output
Cur
___________
Security Code
code
000
___________
Custom Curve
SEtUP
CuSt
___________
Reference temperature
t ref
25.0°C
___________
Calibrate Menu
Cell constant
CELL Const
1.00
___________
Temperature slope
tEMP slOpe
2.000
___________
Output Calibration
OUtPUt CAL
___________
diagnose Menu
diagnose
(Each segment displays
SAMple ReAdINGS
the current value in the transmitter.)
Absolute conductivity
Abs
1000 µS
___________
Off Set
0 AIR
0.0 µS
___________
Cell constant (Linear mode)
CELL COnSt
3.00/cm
___________
Calibration factor
CAL F
1.000
___________
Temperature slope
tslope
2.000
___________
Software version
soft
A02.09
___________
Hardware version
HArd
01
___________
Show fault warnings
FaULTS
none
___________
TAble 1. 5081C-HT default Settings