Emerson Rosemount Analytical HART SMART 5081C-HT Instruction Sheet Download Page 2

2

5081C-HT

Instruction Sheet

January 2015

LIQ_MAN_ABR_5081C-HT/Rev.G

Specifications

pHySICAl SpeCIFICATIoNS

Housing:

Cast  aluminum  with  epoxy  coating.  Type  4X

(IP65). Neoprene O-ring cover seals. 160.5 mm  x 175.3 mm
x 161.3 mm (6.3 in. x 6.9 in. x 6.4 in.)

diameter: 

155.4 mm (6.1 in.) 

electrical Conduit openings:

3/4 in. FNPT 

power  Supply  and  load:

A  minimum  loop  resistance

(load)  of  250  ohms  and  a  minimum  power  supply  of  18
volts DC is required for digital communication as shown in
the load/power supply graph below. 

local Readout:

Main Display is 4 digits, 20 mm tall (0.8 in.)      
Message Display is ten digits, 7 mm tall (0.3 in.)

Automatic Temperture Compensation 

3-wire Pt100 RTD                                      

Conductivity: 0 to 200 °C (32 to 392 °F)                                  
% concentration: 0 to 100 °C (32 to 212°F)
Low Conductivity: 0 to 100 °C (32 to 212 °F)        

Ambient temperature:

-20 to 65° C (-4 to 149° F)

Relative Humidity:

0-95% with enclosure sealed.

Ce: eMI/RFI certified:

EN61326                                                     

HAzARdoUS AReA ClASSIFICATIoN:
Intrinsic Safety:

Class I, II, III,  Div. 1 
Groups A-G
T4  Tamb = 70°C, = 62°C (-FF)

Exia Entity
Class I, Groups A-D
Class II, Groups E-G
Class III
T4

Tamb = 70°C

IECEx BAS 09.0159X
Ex ia IIC T4 Ga

0600  II  1  G

Baseefa03ATEX0099
EEx  ia  IIC  T4
Tamb = -20°C to +65°C

Non-Incendive: 

Class I, Div. 2, Groups A-D
Dust Ignition Proof
Class II & III, Div. 1, Groups E-G
Type 4X Enclosure

Class I, Div. 2, Groups A-D
Suitable for Class II, Div. 2, Groups E-G
T4

Tamb = 70°C

explosion-proof: 

Class I, Div. 1, Groups B-D
Class II, Div. 1, Groups E-G
Class III, Div. 1

Class I, Groups B-D
Class II, Groups E-G
Class III
Tamb = 65°C  max

TRANSMITTeR SpeCIFICATIoNS @ 25°C

Measured Range*

: 0-20,000 µS/cm

Accuracy:

± 0.5% of reading and ± 0.001 µS/cm

Repeatability:

± 0.25% of reading

Stability:

0.25% of output range/month, non-cumulative

Ambient Temperature Coefficient:

± 0.05% of reading/°C

Temperature Slope Adjustment:

0-5%/° C

Other temperature compensation algorithms: ultra-pure
water compensation, cation conductivity, or raw (uncom-
pensated) conductivity.

Compatible  RTd: 

100

W

or  1000

W

with  Automatic

Recognition

loop SpeCIFICATIoNS

loop  Accuracy:

Under  controlled  laboratory  conditions

at 25°C (77°F) with perfectly calibrated ENDURANCE sen-
sor of appropriate cell constant:

up to 5,000 µS/cm:

± 1.0% and ± 2 least significant digit

from 5,000 to 20,000:

± 2% of reading and ± 2 least sig-

nificant digit

SeNSoR SeleCTIoN GUIdelINeS

Note: 

The conductivity values shown in the above chart are for UNCOMPENSATED (or

RAW) conductivity at 25°C.  Maximum range values will vary due to temperature com-

pensation selection, process temperature, and other process conditions.

ATEX

Cell Constant

Suggested Conductivity Range

0.01/cm

up to 50 µS/cm

0.1/cm

1.0 to 500 µS/cm

1.0/cm

10 to 20,000 µS/cm

Summary of Contents for Rosemount Analytical HART SMART 5081C-HT

Page 1: ...d maintaining the instrument To ensure their continued operation to the design specifications personnel should read this manual thoroughly before proceeding with installation commissioning operation and maintenance of this instrument If this equipment is used in a manner not specified by the manufacturer the protection provided by it against hazards may be impaired Failure to follow the proper ins...

Page 2: ...D Dust Ignition Proof Class II III Div 1 Groups E G Type 4X Enclosure Class I Div 2 Groups A D Suitable for Class II Div 2 Groups E G T4 Tamb 70 C explosion proof Class I Div 1 Groups B D Class II Div 1 Groups E G Class III Div 1 Class I Groups B D Class II Groups E G Class III Tamb 65 C max TRANSMITTeR SpeCIFICATIoNS 25 C Measured Range 0 20 000 µS cm Accuracy 0 5 of reading and 0 001 µS cm Repea...

Page 3: ...RESET to end the current operation and return to the main display Changes will NOT be saved ReSeT does NoT return the transmitter to factory default settings CAl Press CAL to access the calibration menu pRoG Press PROG to access the program menu dIAG Press DIAG to read diagnostic mes sages Hold Press HOLD to access the prompts used for turning on or off the hold function ARRoW KeyS Use é and ê key...

Page 4: ...OutPut OUtPUt 1 Not available with display in Neutral Salt Cation or Resistivity modes OUtPUt Current output menu header 4MA 4mA current output setpoint 20MA 20mA current output setpoint HoLd Current output on hold FAULt Fault condition current output setting dPn Current output dampening time tESt Current output test value tEMP Temperature menu header tAUtO Automatic temperature compensation tMAn ...

Page 5: ...smitter measurement and configuration variables The user can read raw data final data and program settings FIGURe 3 AMS configuration screen using HART and can reconfigure the transmitter from anywhere in the plant The figures below show two of the many configura tion and measurement screens available using HART AMS FIGURe 4 AMS measurement screen using HART ...

Page 6: ...Cust Temperature C or F temp C ___________ Output mA or output Cur ___________ Security Code code 000 ___________ Custom Curve SEtUP CuSt ___________ Reference temperature t ref 25 0 C ___________ Calibrate Menu Cell constant CELL Const 1 00 ___________ Temperature slope tEMP slOpe 2 000 ___________ Output Calibration OUtPUt CAL ___________ diagnose Menu diagnose Each segment displays SAMple ReAdI...

Page 7: ...actory terminated PN 23747 00 and unterminated PN 9200275 connecting cable are available The use of fac tory terminated cable is strongly recommended To pre pare unterminated cable for use follow the instructions in the sensor instruction manual Recommended Transmitter Settings Assumes 1 0 cell constant Software Setting process Stream Type Contacting Conductivity in the range 20 to LInEAr 20 000 m...

Page 8: ... the ribbon cable between the display board and the CPU board Rotate the display The black infrared sensor will be at the top of the display INSTAllATIoN See Figure 7 1 Although the analyzer is suitable for outdoor use do not install it in direct sunlight or in areas of extreme temperatures 2 Install the analyzer in an area where vibrations and electromagnetic and radio frequency interference are ...

Page 9: ... 00 F ROSEMOUNT ANALYTICAL Tamb 20 C TO 65 C I S SUPPLY Ui 30 VDC Ii 200 mA Pi 0 9 W Ci 0 µF Li 0 µH SIGNAL INPUT Uo 6 51 V Io 400 mA Po 642 mW Ci 24 1 µF Li 0 µH 1180 MODEL 5081 C HT 73 IECExBAS09 0159X Baseefa03ATEX0099X Ex ia IIC T4 Ga II 1 G FIGURe 8 ATeX label Information ...

Page 10: ...10 5081C HT Instruction Sheet January 2015 LIQ_MAN_ABR_5081C HT Rev G Figure 9 FMRC explosion proof Installation ...

Page 11: ...11 Instruction Sheet 5081C HT LIQ_MAN_ABR_5081C HT Rev G January 2015 Figure 10 FM Hazardous Area Approval label ...

Page 12: ...BSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR SUITABILITY FOR DIVISION 2 13 14 12 13 UNCLASSIFIED AREA D2 12 DIVISION 2 WIRING METHOD PER THE NEC EXCLUDING NONINCENDIVE FIELD WIRING 13 INSTALATION TO BE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE D1 OUTPUT PARAMETERS It Vt TABLE II MODEL 5081 C HT TB1 1 THRU 12 Ci µ F Li mH 30 300 1 0 MODEL NO 275 Voc max OUT Vdc Isc max OUT mA ENTIT...

Page 13: ...13 Instruction Sheet 5081C HT LIQ_MAN_ABR_5081C HT Rev G January 2015 Figure 12 CSA Hazardous Area Approval label ...

Page 14: ...14 5081C HT Instruction Sheet January 2015 LIQ_MAN_ABR_5081C HT Rev G Figure 13 CSA Intrinsically Safe Installation ...

Page 15: ...ual temperature use the arrow keys then press ENTER 11 If you selected CondUC or rStvty in step 2 you will see a CoMp Temperature Compensation type screen Use the arrow keys to select desired temperature compensation lineAr linear n SAlt neutral salt CatIon cation conductivity or none raw or uncompensated conductivity Press ENTER If you are in lineAR mode you can now enter a particular temperature...

Page 16: ...marks are the property of their respective owners The contents of this publication are presented for information purposes only and while effort has been made to ensure their accuracy they are not to be construed as warranties or guarantees express or implied regarding the products or services described herein or their use or applicability All sales are governed by our terms and conditions which ar...

Reviews: