15
Instruction Sheet
5081C-HT
LIQ_MAN_ABR_5081C-HT/Rev.G
January 2015
Engineering Specifications
qUICK-START
1. On the Remote, press PROG, NEXT, NEXT, ENTER.
2. Use the arrow buttons to select
Conduc
(conductivity),
rStvty
(resistivity), or
CuSt
(custom
curve) mode. Press ENTER. If you chose
ConduC
or
rStvty
, go to step 4.
3.
If you selected CuST
, you will see the Setup Custom
screen. To move to the custom curve configuration
menu, press ENTER. You will automatically return to
this same Setup Custom screen after configuration is
complete. To continue transmitter display
programming, press NEXT in the Setup Custom screen.
4. Use the arrow keys to toggle temperature units
between Celsius and Farenheit.
5. Press ENTER then RESET.
6. Press PROG, ENTER.
7. Use the arrow buttons to enter the 4 mA value. Press
ENTER.
8. Use the arrow buttons to enter the 20 mA value.
Press ENTER then RESET.
9. Press PROG, NEXT, ENTER.
10. Use the arrow key to toggle
tAuto
to On or OFF to
select using either the process temperature (
tAuto
=
On) or a manual temperature (
tAuto
= OFF). Press
ENTER. If you selected
tAuto
= OFF, you will be
prompted to enter the manual temperature; use the
arrow keys, then press ENTER.
11. If you selected
CondUC
or
rStvty
in step 2, you will
see a
CoMp
(Temperature Compensation type)
screen. Use the arrow keys to select desired
temperature compensation:
lineAr
(linear),
n SAlt
(neutral salt),
CatIon
(cation conductivity), or
none
(raw or uncompensated conductivity). Press ENTER. If
you are in
lineAR
mode, you can now enter a
particular temperature slope (default is 2%/degC),
then press ENTER to apply the slope.
12. Press RESET.
13. Press CAL, NEXT, NEXT, NEXT, ENTER.
14. Use the arrow buttons to enter the cell constant of
the sensor. Press ENTER, then EXIT.
15. If you are in
rStvty
,
n SAlt
,
Cation
, or
none
(raw)
mode, quick start is complete, proceed to step 20.
16. If you are in
lineAR
mode, press EXIT, CAL, NEXT,
ENTER.
17. Hold the sensor in air to zero. Press ENTER, then EXIT.
18. Place the sensor in a solution of known conductivity
value. Press CAL, ENTER.
19. Use the arrow buttons to enter the current
conductivity value of the solution. Press ENTER.
20. Press RESET.
To ReSeT TRANSMITeR To FACToRy deFAUlT
SeTTINGS:
1. Press PROGRAM, NEXT, NEXT, NEXT, NEXT, NEXT. The
screen should say
“deFAUlt”
. Press ENTER.
2. Use the arrow keys to toggle between
no
(retain your
configuration and calibration settings) and
yeS
(restore factory default settings to all variables).
3. Press ENTER, then EXIT.