Instruction Sheet

5081C-HT

LIQ_MAN_ABR_5081C-HT/Rev.G

January 2015

Engineering Specifications

qUICK-START

1.  On the Remote, press PROG, NEXT, NEXT, ENTER.
2.  Use the arrow buttons to select 

Conduc

(conductivity), 

rStvty

(resistivity), or

CuSt

(custom

curve) mode. Press ENTER. If you chose 

ConduC

or

rStvty

, go to step 4.

3. 

If you selected CuST

, you will see the Setup Custom

screen. To move to the custom curve configuration

menu, press ENTER. You will automatically return to

this same Setup Custom screen after configuration is

complete. To continue transmitter display

programming, press NEXT in the Setup Custom screen.

4.  Use the arrow keys to toggle temperature units

between Celsius and Farenheit.

5.  Press ENTER then RESET.
6.  Press PROG, ENTER.
7.  Use the arrow buttons to enter the 4 mA value. Press

ENTER.

8.  Use the arrow buttons to enter the 20 mA value.

Press ENTER then RESET.

9.  Press PROG, NEXT, ENTER.
10. Use the arrow key to toggle 

tAuto

to On or OFF to

select using either the process temperature (

tAuto

=

On) or a manual temperature (

tAuto

= OFF). Press

ENTER. If you selected 

tAuto

= OFF, you will be

prompted to enter the manual temperature; use the

arrow keys, then press ENTER.

11. If you selected 

CondUC

or 

rStvty

in step 2, you will

see a 

CoMp

(Temperature Compensation type)

screen. Use the arrow keys to select desired

temperature compensation: 

lineAr

(linear), 

n SAlt

(neutral salt), 

CatIon

(cation conductivity), or 

none

(raw or uncompensated conductivity). Press ENTER. If

you are in 

lineAR

mode, you can now enter a

particular temperature slope (default is 2%/degC),

then press ENTER to apply the slope.

12. Press RESET.
13. Press CAL, NEXT, NEXT, NEXT, ENTER.
14. Use the arrow buttons to enter the cell constant of

the sensor. Press ENTER, then EXIT.

15. If you are in 

rStvty

n SAlt

Cation

, or 

none

(raw)

mode, quick start is complete, proceed to step 20.

16. If you are in 

lineAR

mode, press EXIT, CAL, NEXT,

ENTER.

17. Hold the sensor in air to zero. Press ENTER, then EXIT.
18. Place the sensor in a solution of known conductivity

value. Press CAL, ENTER.

19. Use the arrow buttons to enter the current

conductivity value of the solution. Press ENTER.

20. Press RESET.

To ReSeT TRANSMITeR To FACToRy deFAUlT

SeTTINGS:

1.  Press PROGRAM, NEXT, NEXT, NEXT, NEXT, NEXT. The

screen should say 

“deFAUlt”

. Press ENTER.

2.  Use the arrow keys to toggle between 

no

(retain your

configuration and calibration settings) and 

yeS

(restore factory default settings to all variables). 

3.  Press ENTER, then EXIT.

Summary of Contents for Rosemount Analytical HART SMART 5081C-HT

Page 1: ...d maintaining the instrument To ensure their continued operation to the design specifications personnel should read this manual thoroughly before proceeding with installation commissioning operation and maintenance of this instrument If this equipment is used in a manner not specified by the manufacturer the protection provided by it against hazards may be impaired Failure to follow the proper ins...

Page 2: ...D Dust Ignition Proof Class II III Div 1 Groups E G Type 4X Enclosure Class I Div 2 Groups A D Suitable for Class II Div 2 Groups E G T4 Tamb 70 C explosion proof Class I Div 1 Groups B D Class II Div 1 Groups E G Class III Div 1 Class I Groups B D Class II Groups E G Class III Tamb 65 C max TRANSMITTeR SpeCIFICATIoNS 25 C Measured Range 0 20 000 µS cm Accuracy 0 5 of reading and 0 001 µS cm Repea...

Page 3: ...RESET to end the current operation and return to the main display Changes will NOT be saved ReSeT does NoT return the transmitter to factory default settings CAl Press CAL to access the calibration menu pRoG Press PROG to access the program menu dIAG Press DIAG to read diagnostic mes sages Hold Press HOLD to access the prompts used for turning on or off the hold function ARRoW KeyS Use é and ê key...

Page 4: ...OutPut OUtPUt 1 Not available with display in Neutral Salt Cation or Resistivity modes OUtPUt Current output menu header 4MA 4mA current output setpoint 20MA 20mA current output setpoint HoLd Current output on hold FAULt Fault condition current output setting dPn Current output dampening time tESt Current output test value tEMP Temperature menu header tAUtO Automatic temperature compensation tMAn ...

Page 5: ...smitter measurement and configuration variables The user can read raw data final data and program settings FIGURe 3 AMS configuration screen using HART and can reconfigure the transmitter from anywhere in the plant The figures below show two of the many configura tion and measurement screens available using HART AMS FIGURe 4 AMS measurement screen using HART ...

Page 6: ...Cust Temperature C or F temp C ___________ Output mA or output Cur ___________ Security Code code 000 ___________ Custom Curve SEtUP CuSt ___________ Reference temperature t ref 25 0 C ___________ Calibrate Menu Cell constant CELL Const 1 00 ___________ Temperature slope tEMP slOpe 2 000 ___________ Output Calibration OUtPUt CAL ___________ diagnose Menu diagnose Each segment displays SAMple ReAdI...

Page 7: ...actory terminated PN 23747 00 and unterminated PN 9200275 connecting cable are available The use of fac tory terminated cable is strongly recommended To pre pare unterminated cable for use follow the instructions in the sensor instruction manual Recommended Transmitter Settings Assumes 1 0 cell constant Software Setting process Stream Type Contacting Conductivity in the range 20 to LInEAr 20 000 m...

Page 8: ... the ribbon cable between the display board and the CPU board Rotate the display The black infrared sensor will be at the top of the display INSTAllATIoN See Figure 7 1 Although the analyzer is suitable for outdoor use do not install it in direct sunlight or in areas of extreme temperatures 2 Install the analyzer in an area where vibrations and electromagnetic and radio frequency interference are ...

Page 9: ... 00 F ROSEMOUNT ANALYTICAL Tamb 20 C TO 65 C I S SUPPLY Ui 30 VDC Ii 200 mA Pi 0 9 W Ci 0 µF Li 0 µH SIGNAL INPUT Uo 6 51 V Io 400 mA Po 642 mW Ci 24 1 µF Li 0 µH 1180 MODEL 5081 C HT 73 IECExBAS09 0159X Baseefa03ATEX0099X Ex ia IIC T4 Ga II 1 G FIGURe 8 ATeX label Information ...

Page 10: ...10 5081C HT Instruction Sheet January 2015 LIQ_MAN_ABR_5081C HT Rev G Figure 9 FMRC explosion proof Installation ...

Page 11: ...11 Instruction Sheet 5081C HT LIQ_MAN_ABR_5081C HT Rev G January 2015 Figure 10 FM Hazardous Area Approval label ...

Page 12: ...BSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR SUITABILITY FOR DIVISION 2 13 14 12 13 UNCLASSIFIED AREA D2 12 DIVISION 2 WIRING METHOD PER THE NEC EXCLUDING NONINCENDIVE FIELD WIRING 13 INSTALATION TO BE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE D1 OUTPUT PARAMETERS It Vt TABLE II MODEL 5081 C HT TB1 1 THRU 12 Ci µ F Li mH 30 300 1 0 MODEL NO 275 Voc max OUT Vdc Isc max OUT mA ENTIT...

Page 13: ...13 Instruction Sheet 5081C HT LIQ_MAN_ABR_5081C HT Rev G January 2015 Figure 12 CSA Hazardous Area Approval label ...

Page 14: ...14 5081C HT Instruction Sheet January 2015 LIQ_MAN_ABR_5081C HT Rev G Figure 13 CSA Intrinsically Safe Installation ...

Page 15: ...ual temperature use the arrow keys then press ENTER 11 If you selected CondUC or rStvty in step 2 you will see a CoMp Temperature Compensation type screen Use the arrow keys to select desired temperature compensation lineAr linear n SAlt neutral salt CatIon cation conductivity or none raw or uncompensated conductivity Press ENTER If you are in lineAR mode you can now enter a particular temperature...

Page 16: ...marks are the property of their respective owners The contents of this publication are presented for information purposes only and while effort has been made to ensure their accuracy they are not to be construed as warranties or guarantees express or implied regarding the products or services described herein or their use or applicability All sales are governed by our terms and conditions which ar...

Reviews: