Quick Installation Guide

00825-0100-4662, Rev CC
December 2012

Rosemount 8732

Start

Step 1: Pre-Installation

Step 2: Handling

Step 3: Mounting

Step 4: Installation 

(Flanged Sensors)

(Wafer Sensors)

(Sanitary Sensors)

Step 5: Grounding

Step 6: Wiring

Step 7: Basic Configuration

Product Certifications

End

Rosemount 8732E Magnetic Flowmeter System 
(Transmitter and Sensor)

4662RevCCQIG.fm  Page 1  Thursday, January 10, 2013  5:11 PM

Summary of Contents for Rosemount 8732E

Page 1: ...2 Rosemount 8732 Start Step 1 Pre Installation Step 2 Handling Step 3 Mounting Step 4 Installation Flanged Sensors Wafer Sensors Sanitary Sensors Step 5 Grounding Step 6 Wiring Step 7 Basic Configuration Product Certifications End Rosemount 8732E Magnetic Flowmeter System Transmitter and Sensor ...

Page 2: ... anything through the sensor for the purpose of lifting or gaining leverage Liner damage can render the sensor useless To avoid possible damage to the sensor liner ends do not use metallic or spiral wound gaskets If frequent removal is anticipated take precautions to protect the liner ends Short spool pieces attached to the sensor ends are often used for protection Correct flange bolt tightening i...

Page 3: ... requirements Mechanical Considerations The mounting site for the Rosemount 8732 transmitter should provide enough room for secure mounting easy access to conduit ports full opening of the transmitter covers and easy readability of the LOI screen see Figure 1 If the Rosemount 8732 is mounted separately from the sensor it is not subject to limitations that might apply to the sensor NOTE M20 and PG ...

Page 4: ...ardware 2 Attach the Rosemount 8732 to the mounting bracket using the mounting screws Identify Options and Configurations The standard application of the 8732 includes a 4 20 mA output and control of the sensor coils and electrodes Other applications may require one or more of the following configurations or options Multidrop Communications Digital Output Digital Input Pulse Output Additional opti...

Page 5: ...rical connections to the Rosemount 8732 consider local and plant electrical standards and be sure to have the proper power supply conduit and other accessories necessary to comply with these standards Rotate Transmitter Housing The electronics housing can be rotated on the sensor in 90 increments by loosening the four mounting bolts on the bottom of the housing and reinstalling the bolts When the ...

Page 6: ...e transport the system to the installation site in the original shipping containers PTFE lined sensors are shipped with end covers that protect it from both mechanical damage and normal unrestrained distortion Remove the end covers just before installation Figure 2 Rosemount 8705 Sensor Support for Handling through 4 Inch Sensors 6 Inch and Larger Sensors ...

Page 7: ...de plane see Figure 3 Installations with reduced straight runs from 0 to five pipe diameters are possible In reduced straight pipe run installations performance will shift to as much as 0 5 of rate Reported flow rates will still be highly repeatable Flow Direction The sensor should be mounted so the FORWARD end of the flow arrow shown on the sensor identification tag points in the direction of flo...

Page 8: ...orizontal installation should be restricted to low piping sections that are normally full Sensor Orientation The electrodes in the sensor are properly orientated when the two measurement electrodes are in the 3 and 9 o clock positions or within 45 from the vertical as shown on the right of Figure 5 Avoid any mounting orientation that positions the top of the sensor at 90 from the vertical position...

Page 9: ...eam side and then snug and tighten the downstream side Failure to alternate between the upstream and downstream flanges when tightening bolts may result in liner damage Suggested torque values by sensor line size and liner type are listed in Table 1 for ASME B16 5 flanges and Table 2 for EN flanges Consult the factory if the flange rating of the sensor is not listed Tighten flange bolts on the ups...

Page 10: ...nge Bolt Torquing Sequence Table 1 Suggested Flange Bolt Torque Values for Rosemount 8705 and 8707 High Signal Sensors PTFE ETFE PFA liners Polyurethane Neoprene Linatex Adiprene liner Size Code Line Size Class 150 pound feet Class 300 pound feet Class 150 pound feet Class 300 pound feet 005 0 5 inch 15 mm 8 8 010 1 inch 25 mm 8 12 015 1 5 inch 40 mm 13 25 7 18 020 2 inch 50 mm 19 17 14 11 025 2 5...

Page 11: ... 5 inch 15 mm 10 010 1 inch 25 mm 20 015 1 5 inch 40 mm 50 020 2 inch 50 mm 60 025 2 5 inch 65 mm 50 030 3 inch 80 mm 50 040 4 inch 100 mm 50 70 050 5 0 inch 125 mm 70 100 060 6 inch 150mm 90 130 080 8 inch 200 mm 130 90 130 170 100 10 inch 250 mm 100 130 190 250 120 12 inch 300 mm 120 170 190 270 140 14 inch 350 mm 160 220 320 410 160 16 inch 400 mm 220 280 410 610 180 18 inch 450 mm 190 340 330 ...

Page 12: ...eter Newton meter Newton meter 010 1 inch 25 mm 20 015 1 5 inch 40 mm 30 020 2 inch 50 mm 40 025 2 5 inch 65 mm 35 030 3 inch 80 mm 30 040 4 inch 100 mm 40 50 050 5 0 inch 125 mm 50 70 060 6 inch 150 mm 60 90 080 8 inch 200 mm 90 60 90 110 100 10 inch 250 mm 70 80 130 170 120 12 inch 300 mm 80 110 130 180 140 14 inch 350 mm 110 150 210 280 160 16 inch 400 mm 150 190 280 410 180 18 inch 450 mm 130 ...

Page 13: ...d for the spacers provided Stud specifications are listed in Table 3 3 Place the sensor between the flanges Make sure that the alignment spacers are properly centered on the studs For vertical flow installations slide the o ring over the stud to keep the spacer in place See Figure 8 To ensure the spacers match the flange size and class rating for the process flanges see Table 4 4 Insert the remain...

Page 14: ... ANSI 150 AA80 8 200 ANSI 150 AB15 1 5 40 ANSI 300 AB20 2 50 ANSI 300 AB30 3 80 ANSI 300 AB40 4 100 ANSI 300 AB60 6 150 ANSI 300 AB80 8 200 ANSI 300 AB15 1 5 40 ANSI 300 AB20 2 50 ANSI 300 AB30 3 80 ANSI 300 AB40 4 100 ANSI 300 AB60 6 150 ANSI 300 AB80 8 200 ANSI 300 DB40 4 100 DIN PN10 16 DB60 6 150 DIN PN10 16 DB80 8 200 DIN PN10 16 DC80 8 100 DIN PN25 DD15 1 5 150 DIN PN10 16 25 40 DD20 2 50 DI...

Page 15: ...ess connection fitting such as a Tri Clamp fitting on all Rosemount 8721 Sanitary sensors except when the process connection fittings are not supplied and the only connection type is an IDF fitting Alignment and Bolting Standard plant practices should be followed when installing a magmeter with sanitary fittings Unique torque values and bolting techniques are not required Table 5 Rosemount 8711 To...

Page 16: ...e 6 Process Grounding Installation Process Grounding Options Type of Pipe Grounding Straps Grounding Rings Grounding Electrode Lining Protectors Conductive Unlined Pipe See Figure 10 Not Required Not Required See Figure 11 Conductive Lined Pipe Insufficient Grounding See Figure 11 See Figure 10 See Figure 11 Non Conductive Pipe Insufficient Grounding See Figure 12 See Figure 13 See Figure 12 Figur...

Page 17: ...llation Guide 00825 0100 4662 Rev CC December 2012 Rosemount 8732 17 Figure 12 Grounding with Grounding Rings or Lining Protectors Figure 13 Grounding with Grounding Electrode Grounding Rings or Lining Protectors ...

Page 18: ... dedicated conduit line between each transmitter and sensor is required Shielded cable must be used for best results in electrically noisy environments When preparing all wire connections remove only the insulation required to fit the wire completely under the terminal connection Removal of excessive insulation may result in an unwanted electrical short to the transmitter housing or other wire con...

Page 19: ...smitter can be integral to the sensor or remotely mounted following the wiring instructions Remote Mount Cable Requirements and Preparation For installations using the individual coil drive and signal cable lengths should be limited to less than 1 000 feet 300 meters Equal length cable is required for each See Table 7 For installations using the combination coil drive and signal cable lengths shou...

Page 20: ...le 9 Combination Coil and Signal Cable Table 7 Cable Requirements Description Length Part Number Coil Drive Cable 14 AWG Belden 8720 Alpha 2442 or equivalent ft m 08712 0060 0001 08712 0060 2013 Signal Cable 20 AWG Belden 8762 Alpha 2411 or equivalent ft m 08712 0061 0001 08712 0061 2003 Combination Cable Coil Drive Cable 18 AWG and Signal Cable 20 AWG ft m 08732 0753 1003 08732 0753 2004 WARNING ...

Page 21: ...shield wires should be tied with the main shield wire at terminal 17 at the sensor terminal block and cut back to the insulation in the transmitter junction box See Figure 17 Figure 16 Remote Mount Wiring Diagrams Figure 17 Combination Coil and Signal Cable Wiring Diagram Transsmiter Tube Coil Drive Cable 1 Red 2 Green 3 Shield 17 Shield 18 Black 19 White Cut Shield Signal Cable 17 Shield 18 Black...

Page 22: ...P 6 CONTINUED Integral Mount Transmitters The interconnecting wire harness for an integral mount transmitter is installed at the factory See Figure 18 Do not use cable other than that supplied by Emerson Process Management Rosemount Inc Figure 18 8732EST Integral Mount Wiring Diagram ...

Page 23: ...ignal may be powered internally or externally The default position of this switch is in the internal position and is located on the front of the electronics board 8732E terminal connections Connect negative DC to Terminal 1 and positive DC to Terminal 2 See Figure 19 Internal Power Source The 4 20 mA analog signal loop is powered from the transmitter itself External Power Source The 4 20 mA analog...

Page 24: ...itters with extended cable lengths verify that there is a minimum of 12 V DC at the terminals of the transmitter Disconnects Connect the device through an external disconnect or circuit breaker Clearly label the disconnect or circuit breaker and locate it near the transmitter and per local electrical control Installation Category The installation category for the 8732E is Overvoltage Category II O...

Page 25: ...transmitter housings shipped with a cover jam screw the screw should be properly installed once the transmitter has been wired and powered up Follow these steps to install the cover jam screw 1 Verify that the cover jam screw is completely threaded into the housing 2 Install the transmitter housing cover and verify that the cover is tight against the housing 3 Using an M4 hex wrench loosen the jam...

Page 26: ...ld be selected to meet your particular metering needs Line Size The line size sensor size must be set to match the actual sensor connected to the transmitter The size must be specified in inches URV Upper Range Value The upper range value URV sets the 20 mA point for the analog output This value is typically set to full scale flow The units that appear will be the same as those selected under the ...

Page 27: ...of the display To deactivate the display lock DL hold the UP arrow for 10 seconds Once deactivated the DL will no longer appear in the lower right hand corner of the display Function HART Fast Keys Process Variables 1 1 Primary Variable PV 1 1 1 PV Percent of Range 1 1 2 PV Analog Output AO 1 1 3 Totalizer Set Up 1 1 4 Totalizer Units 1 1 4 1 Gross Total 1 1 4 2 Net Total 1 1 4 3 Reverse Total 1 1...

Page 28: ... SNR Signal Power 8714i R esults D A T rim Digital T rim A uto Zero Universal T rim 4 20 mA V erify V iew R esults T ag Flow Units L ine Size PV UR V PV L R V Cal Number PV Damping Coil Frequency Proc Density PV L SL PV USL PV Min Span A nalog Pulse DI DO Config T otalizer R everse Flow HA R T PV UR V PV L R V A larm T ype T est Pulse Scaling Pulse Width Pulse Mode T est DI 1 DO 2 T otalize Units ...

Page 29: ...te EEx e or EEx n metal cable gland and metal blanking plug or any appropriate ATEX approved cable gland and blanking plug with IP66 rating certified by an EU approved certification body CE Marking Complies with EN 61326 1 2006 For Rosemount 8732E transmitters Complies with Essential Health and Safety Requirements EN 60079 0 2006 EN 60079 1 2007 EN 60079 7 2007 EN 60079 11 2007 EN 60079 26 2004 EN...

Page 30: ...Requires sensors with N5 Approval E5 Explosion proof for Class I Division 1 Groups C and D T6 at 60 C Dust ignition proof Class II III Division 1 Groups E F and G T5 at 60 C Non incendive for Class I Division 2 Groups A B C and D flammable fluids T4 at 60 C 50 C Ta 60 C Hazardous Locations Enclosure Type 4X Canadian Standards Association CSA N0 Non incendive for Class I Division 2 Groups A B C and...

Page 31: ...s The property class of the security screws which attach the flowtube or junction box to the transmitter is SST A2 70 Installation Instructions The cable and conduit entry devices and blanking elements shall be of a certified flameproof type suitable for the conditions of use and correctly installed With the use of conduit a certified stopping box shall be provided immediately to the entrance of t...

Page 32: ...s of the flameproof joints The property class of the security screws which attach the flowtube or junction box to the transmitter is SST A2 70 Installation Instructions The cable and conduit entry devices and blanking elements shall be of a certified flameproof or increased safety type suitable for the conditions of use and correctly installed With the use of conduit a certified stopping box shall...

Page 33: ...071438X Ex de IIB T6 or Ex de ia IIB T6 without LOI 50 C Ta 60 C with LOI 20 C Ta 60 C Vmax 250 V AC or 42 V DC InMetro Brazil E2 InMetro Flameproof Certificate No NCC 12 1177 X Ex de IIC T6 Gb IP66 or Ex de ia IIC Ga IIC T6 Gb IP66 without LOI 50 C Ta 60 C with LOI 20 C Ta 60 C Vmax 250 V AC or 42 V DC EB InMetro Flameproof Certificate No NCC 12 1177 X Ex de IIB T6 Gb IP66 or Ex de ia IIC Ga IIB ...

Page 34: ...20 C Ta 60 C Vmax 250 V AC or 42 V DC EK KOSHA Flameproof Certificate No 2008 2094 Q1X Ex de IIB or Ex de ia IIB T6 without LOI 50 C Ta 60 C with LOI 20 C Ta 60 C Vmax 250 V AC or 42 V DC GOST Russia E8 GOST Flameproof Ex de IIC T6 or Ex de ia IIC T6 without LOI 50 C Ta 60 C with LOI 20 C Ta 60 C IP67 EM GOST Flameproof Ex de IIB T6 or Ex de ia IIB T6 without LOI 50 C Ta 60 C with LOI 20 C Ta 60 C...

Page 35: ...rtifications are available on the Rosemount 570TM Rosemount 8705 Sensor Rosemount 8707 Sensor Rosemount 8711 Sensor Rosemount 8721 Sensors NA N0 ND N1 N5 N7 NF E1 E2 E3 E5 2 2 Available in line sizes up to 8 in 200 mm only E8 E9 EB EK EM EP KD For Non flammable Fluids For Flammable Fluids For Non flammable Fluids For Non flammable Fluids For Flammable Fluids For Flammable Fluids For Non flammable ...

Page 36: ... Technology Drive and 8200 Market Boulevard Eden Prairie MN 55344 3695 Chanhassen MN 55317 9687 USA USA to which this declaration relates is in conformity with the provisions of the European Community Directives including the latest amendments as shown in the attached schedule Assumption of conformity is based on the application of harmonized or applicable technical standards and when applicable o...

Page 37: ...odel 8732E Magnetic Flowmeter Transmitter KEMA 07ATEX0073 X Flameproof with Increased Safety Terminal s Intrinsically Safe Output s Dust Equipment Group II Category 2 G Ex d IIB IIC T6 Ex de IIB IIC T6 Ex e IIB IIC Junctionbox Equipment Group II Category 2 1 G Ex de ia IIB IIC T6 Transmitter Equipment Group II Category 1 G Ex ia IIC Equipment Group II Category 1 D Ex tD A20 IP66 T100 ºC EN 60079 0...

Page 38: ... II Category 3 1 G Ex nA nL ia IIC T4 EN 60079 0 2006 EN 60079 15 2005 EN 60079 11 2007 ATEX Notified Bodies for EC Type Examination Certificate KEMA Notified Body Number 0344 Utrechtseweg 310 6812 AR Arnhem P O Box 5185 6802 ED Arnhem The Netherlands Postbank 6794687 Baseefa Notified Body Number 1180 Rockhead Business Park Staden Lane Buxton Derbyshire SK17 9RZ United Kingdom ATEX Notified Body f...