![Emerson Rosemount 644 with HART Protocol Reference Manual Download Page 81](http://html.mh-extra.com/html/emerson/rosemount-644-with-hart-protocol/rosemount-644-with-hart-protocol_reference-manual_104732081.webp)
73
Reference Manual
00809-0200-4728, Rev SA
Troubleshooting
July 2018
Troubleshooting
AMS Device Manager
1. Right click on the device and select
Configure
.
2. In the left navigation pane choose
Manual Setup
and choose the
Sensor 1
or
Sensor 2
tab
depending on the need.
3. Find the
Transmitter Sensor Matching (CVD)
group box and enter in the required CVD constants. Or
select “Set CVD Coefficients” button to be guided through steps. You may also select “Show CVD
Coefficients” button to see the current coefficients loaded into the device.
4. Select
Apply
when complete.
Note
When the transmitter-senor matching is disabled, the transmitter reverts to either user or factory trim,
whichever was used previously. Make certain the transmitter engineering units default correctly before
placing the transmitter into service.
5.7
Switching HART Revision
Some systems are not capable of communicating with HART Revision 7 devices. The following
procedures list how to change HART revisions between HART Revision 7 and HART Revision 5.
5.7.1
Generic menu
If the HART configuration tool is not capable of communicating with a HART Revision 7 device, it should
load a Generic Menu with limited capability. The following procedures allow for switching between HART
Revision 7 and HART Revision 5 from a generic menu in any HART compliant configuration tool.
1. Locate “Message” field.
a. To change to HART Revision 5, Enter:
HART5
in the message field.
b. To change to HART Revision 7, Enter:
HART7
in the message field.
5.7.2
Field Communicator
From the
HOME
screen, enter the Fast Key sequence and follow steps within the Field Communicator to
complete the HART revision change.
5.7.3
AMS Device Manager
1. Right click on the device and select
Configure
.
2. In the left navigation pane choose
Manual Setup
and click on the
HART
tab.
3. Select the
Change HART Revision
button and follow the prompts.
Note
HART Revision 7 is only compatible with AMS Device Manager 10.5, and greater. AMS Device Manager
version 10.5 requires a software patch to be compatible.
Device Dashboard Fast Keys
2, 2, 8, 3
Summary of Contents for Rosemount 644 with HART Protocol
Page 2: ......
Page 8: ...vi Reference Manual 00809 0200 4728 Rev SA Title Page July 2018 Title Page ...
Page 48: ...40 Reference Manual 00809 0200 4728 Rev SA Configuration July 2018 Configuration ...
Page 62: ...54 Reference Manual 00809 0200 4728 Rev SA Installation July 2018 Installation ...
Page 111: ......