4.10.2
Block instantiation
The Rosemount 5900C supports the use of function block instantiation. Then the number
of blocks and block types can be defined to match specific application needs. The number
of blocks that can be instantiated is only limited by the amount of memory within the
device and the block types that are supported by the device. Instantiation does not apply
to standard device blocks like the Resource and transducer blocks.
By reading the parameter “FREE_SPACE” in the Resource block you can determine how
many blocks you can instantiate. Each block that you instantiate takes up 4.6% of the
“FREE_SPACE”.
Block instantiation is done by the host control system or configuration tool, but not all
hosts implement this functionality. Please refer to your specific host or configuration tool
manual for more information.
4.10.3
Factory configuration
The following fixed configuration of function blocks is provided:
Table 4-4: Function Blocks for the Rosemount 5900C
Function Block
Index
Default Tag
Available
1600
AI 1600
Fixed
Analog Input
1700
AI 1700
Fixed
Analog Input
1800
AI 1800
Fixed
Analog Input
1900
AI 1900
Fixed
Analog Input
2000
AI 2000
Fixed
Analog Input
2100
AI 2100
Fixed
Analog Output
2200
AO 2200
Default, deletable
Analog Output
2300
AO 2300
Default, deletable
PID
2400
PID 2400
Default, deletable
Control Selector
2500
CSEL 2500
Default, deletable
Output Splitter
2600
OSPL 2600
Default, deletable
Signal Characterizer
2700
CHAR 2700
Default, deletable
Integrator
2800
INTEG 2800
Default, deletable
Arithmetic
2900
ARITH 2900
Default, deletable
Input Selector
3000
ISEL 3000
Default, deletable
(1)
See
for more information.
(2)
See
Reference Manual
Configuration
00809-0100-5901
September 2019
Rosemount 5900C Radar Level Gauge
161
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...