The resulting potentials, measured with a clean platinum electrode and saturated
KCl/AgCL reference electrode, should be within ±20 millivolts of the value shown in
. Note solution temperature to ensure accurate interpretation of results.
The ORP value of saturated quinhydrone solution is not stable over long periods of
time. Therefore, make these standards fresh each time they are used.
Table 3-1: ORP of Saturated Quinhydrone Solution (Millivolts)
pH 4 solution
pH 7 solution
Temp ºC
20
25
30
20
25
30
mV
potential
268
264
260
94
87
80
5. Remove the sensor from the buffer, rinse, and install in the process.
Reference Manual
Calibration
00809-0100-3389
February 2019
Reference Manual
29