4
Maintenance
Rosemount
™
389/389VP sensors require minimum maintenance. Keep the sensors clean
and free of debris and sediment at all times. The frequency of cleaning by wiping or
brushing with a soft cloth or brush is determined by the nature of the solution being
measured. Remove sensors from the process periodically and check them in buffer
solutions.
WARNING
PRESSURE AND TEMPERATURE
Before removing the sensor, be absolutely certain that the process pressure is reduced to
0 psig and the process temperature is lowered to a safe level!
If the sensor will not calibrate, refer to your transmitter reference manual for proper test
procedures. If you determine that the sensor has failed, discard and replace it.
4.1
Clean electrode
If the electrode is coated or dirty, clean as follows:
Procedure
1. Remove the sensor from process.
2. Remove the sensor from process as instructed in
3. Wipe the glass bulb with a soft, clean, lint free cloth or tissue. If this does not
remove the dirt or coating, go to
(detergents clean oil and grease; acids
remove scale.)
4. Wash the glass bulb in a strong detergent solution and rinse it in clean water. If this
does not clean the glass bulb, go to
WARNING
CORROSIVE SUBSTANCE
The solution used during the following step is an acid; handle with care. Follow the
directions of the acid manufacturer. Wear the proper protective equipment. Do not
let the solution come in contact with skin or clothing. If contact with skin is made,
immediately rinse with clean water.
5. Wash the bulb in a dilute 5% hydrochloric acid solution and rinse with clean water.
Soaking the sensor overnight in the acid solution can improve cleaning action.
Replace the sensor if it cannot be cleaned.
6. Following the Warning above, wash the glass bult in dilute 5% hydrochloric acid
solution and then rinse it thoroughly with clean tap water. Replace the sensor if it
cannot be clean. If the glass bulb appears clean, proceed to
7. Buffer calibrate the sensor.
Reference Manual
Maintenance
00809-0100-3389
February 2019
Reference Manual
31
Summary of Contents for Rosemount 389
Page 4: ...Contents Reference Manual February 2019 00809 0100 3389 4 Rosemount 389 389VP ...
Page 26: ...Install Reference Manual February 2019 00809 0100 3389 26 Rosemount 389 389VP ...
Page 30: ...Calibration Reference Manual February 2019 00809 0100 3389 30 Rosemount 389 389VP ...
Page 34: ...Maintenance Reference Manual February 2019 00809 0100 3389 34 Rosemount 389 389VP ...
Page 38: ...Return of Material Reference Manual February 2019 00809 0100 3389 38 Rosemount 389 389VP ...
Page 44: ...China RoHS Table Reference Manual February 2019 00809 0100 3389 44 Rosemount 389 389VP ...