Figure 3-20: Base Enclosure security seals
A. Transmitter Electronics Enclosure
B. Security wire seals
C. Transmitter Electronics endcap security latch
D. Base Enclosure
23. If required, install a security seal wire (maximum wire diameter 0.09 inch; 2.3mm)
on the Transducer Cable Connector to the Transducer Cable Nut by rotating the
Transducer Cable Nut clockwise; compressing the seal on the Transducer Cable
Connector (see
Figure 3-21: Transducer cable nut, chordset and security seal
A. Transducer cable nut
B. Transducer cable chordset
C. Security wire seal
Meter repairs
Maintenance and Troubleshooting Manual
April 2022
00809-0600-3104
74