June 2016
8
Quick Start Guide
Figure 4. Rosemount 3095FC Terminal Block and Wiring Diagram
3.0 Set jumpers and apply power
To prevent unnecessary battery drainage, the Rosemount 3095FC is delivered
with the reset jumper in the OFF position. To apply power to the Rosemount
3095FC:
1. Complete the necessary wiring.
2. Unscrew the front end cap cover (LCD display end).
3. Place the power jumper in the ON position. The jumper is located on the LCD
display (if installed) or at J1 on the Battery Charger Board.
4. Screw the front-end cap cover (LCD display end).
After the Rosemount 3095FC completes start-up diagnostics (RAM and other
internal checks), the optional LCD display shows the date and time to indicate
that the Rosemount 3095FC completed a valid reset sequence.
A. LOI
B. RET
C. RTD
D. CHG
E. 2 or 3 wire RTD
F. Power supply
G. PC
H. Operator interface cable
2 OR 3 WIRE RTD
RTD
LOI
OPERATOR
INTERFACE
CABLE
PC
POWER
SUPPLY
+
+
-
+
RET
CHG
A
B
C
D
E
F
G.
H