Reference Manual
00809-0100-4828, Rev EA
November 2007
2-3
Rosemount 1595
Liquid or Steam in Horizontal Pipes
The electronics should be mounted below the pipe to ensure that gases do
not collect on the transmitter sensing diaphragms.
Figure 2-2. Liquid and Steam in
Horizontal Pipes
Vertical Pipe Installation
Gas in Vertical Pipes
Mount the electronics above the pipe with the instrument lines sloping down.
Figure 2-3. Gas in Vertical Pipes
FLOW
Flow