2.2
Mounting
Mount the transmitter at a high point in the conduit run to prevent moisture from draining
into the transmitter housing.
The Rosemount
™
148 Temperature Transmitter installs:
•
In a connection head or universal head mounted directly on a sensor assembly
•
Apart from a sensor assembly using a universal head
•
To a DIN rail using an optional mounting clip
2.2.1
Mounting a Rosemount 148 to a DIN Rail
To attach a head mount transmitter to a DIN rail, assemble the appropriate rail mounting
kit (part number 00248-1601-0001) to the transmitter as shown in
.
Figure 2-1: Assembling Rail Clip Hardware to a Rosemount 148
C
B
A
A. Mounting hardware
B. Transmitter
C. Rail clip
2.3
Installation
The Rosemount
™
148 Temperature Transmitter can be ordered assembled to a sensor
with the XA option code and thermowell or as a stand-alone unit. If ordered without the
sensor assembly, use the following guidelines when installing the transmitter with an
integral sensor assembly.