C.4.3
Analog alarm limits
High/low alarm value
The high/low alarm current for the analog output when the device enters the alarm mode.
Related information
High/low saturation value
The device will continue to set a current that corresponds with the measurement until
reaching the upper/lower limit (and then freeze).
Related information
C.5
Geometry
C.5.1
Advanced
Calibration offset
Difference between surface distance measured by transmitter and the same distance
measured by, for example, hand-dipping with a measurement tape. A positive Calibration
Offset value will increase the presented level value.
Figure C-9: Calibration Offset
A
C
D
B
A. Positive Calibration Offset value
B. Negative Calibration Offset value
C. Reported level
D. Actual level
Configuration parameters
Reference Manual
September 2020
00809-0100-4480
92
Rosemount 1408H Level Transmitter