2
Transmitter overview
2.1
Measurement principle
The Rosemount 1408H is a transmitter for continuous level measurements in the hygienic
industry. The measurement principle is fast-sweep Frequency Modulated Continuous
Wave (FMCW).
The transmitter continuously emits signal sweeps with a constantly varying frequency
towards the product surface. Since the transmitter continuously changes the frequency of
the transmitted signal, there will be a difference in frequency between the transmitted
and the reflected signals (see
The frequency of the reflected signal is subtracted from the frequency of the signal
transmitted at that moment, resulting in a low frequency signal which is proportional to
the distance to the product surface. This signal is further processed to obtain fast, reliable,
and highly accurate level measurements.
Figure 2-1: FMCW-method
d
f
in
f
max
f
min
f
out
f
in
f
out
t
f
A
C
D
B
Δf
∼
d=distance
A. Frequency (GHz)
B. Time (s)
C. Transmitted signal
D. Reflected signal
2.2
Process characteristics
2.2.1
Dielectric constant
A key parameter for measurement performance is reflectivity. A high dielectric constant of
the media provides better reflection and enables a longer measuring range.