Rosemount 1153 Series B
Figure 3-2. Zero and Span
Adjustment Screws.
Example (for Range Code 4)
Initial transmitter calibration: 25 to 125 inH
2
O
(100 inH
2
O span with zero suppressed 25 inH
2
O)
Desired transmitter calibration: –75 to –25 inH
2
O
(50 inH
2
O span with zero elevated 75 inH
2
O)
Zero
Span