5.4.3
Procedure - Configure Outputs
Under the Program/Outputs menu, the adjacent screen appears to
allow configuration of the outputs. Follow the menu screens in
Figure 5-2
to
configure the outputs.
5.4.4
Procedure - Assigning Measurements the Low and
High Current Outputs
The adjacent screen appears when entering the
Assign
function under
Program/Output/Configure. These screens allow you to assign a
measurement, process value, or temperature input to each output.
Follow the menu screens in
Figure 5-2
to assign measurements to
the outputs.
5.4.5
Procedure - Ranging the Current Outputs
The adjacent screen appears under Program/Output/Range.
Enter a value for 4mA and 20mA (or 0 mA and 20 mA) for each output.
Follow the menu screens in
Figure 5-2
to assign values to the outputs.
S1: 1.234µS/cm 123.4°C
S2: 12.34pH 123.4°C
S3:123.4µS/cm 123.4°C
Output
M
Assign
S1 Measurement
S1 Temperature
S2 Measurement
S2 Temperature
S3 Measurement
S3 Temperature
S1: 1.234µS/cm 123.4°C
S2: 12.34pH 123.4°C
S3:123.4µS/cm 123.4°C
Output Range
OM SN 4mA: 0.000µS/cm
OM SN 20mA: 20.00µS/cm
OM SN 4mA: 00.00pH
OM SN 20mA: 14.00pH
S1: 1.234µS/cm 123.4°C
S2: 12.34pH 123.4°C
S3:123.4µS/cm 123.4°C
Output
M
Configure
Assign: S1 Meas
Range: 4-20mA
Scale: Linear
Dampening: 0sec
Fault Mode: Fixed
Fault Value: 21.00mA
Figure 5-2 Configuring and Ranging the Current Outputs
Instruction Manual
LIQ-MAN-1057
Section 5: Programming the Analyzer - Basics
April 2017
Section 5: Programming the Analyzer - Basics
23