MODEL 1054B pH/ORP
SECTION 4.0
CONFIGURATION
RANGE
FACTORY SET
USER SET
A. Alarm 1 Setup
(
AL1
)
1. Alarm Status (
On/OFF
)
On
_________
2. High or Low (
H-L
) (
H i
/
Lo
)
Lo
_________
3. Hysteresis (
HYS
)
0-25% of setpoint
0.0%
_________
4. Delay Time On (
don
)
0-255 sec.
000 Seconds
_________
5. Delay Time Off (
doF
)
0-255 sec.
000 Seconds
_________
B. Alarm 2 Setup
(
AL2
)
1. Alarm Status (
On
/
FLt
/
OFF
)
On
_________
2. High or Low (
H-L
) (
H i
/
Lo
)
Hi
_________
3. Hysteresis (
HYS
)
0-25% of setpoint
0.0%
_________
4. Delay Time On (
don
)
0-255 sec.
000 Seconds
_________
5. Delay Time Off (
doF
)
0-255 sec.
000 Seconds
_________
C. Interval Timer
(
Int
)
1. Active Status (
tOn
) (
oFF
/
on
)
oFF
_________
2. Interval Time (
int
)
minimum 10 minutes
1 Day
_________
3. Count (
Cnt
)
1 to 60
5
_________
4. On Time (
Ont
)
0 to 299.9 sec
1 Second
_________
5. Off Time (
OFt
)
0 to 299.9 sec
1 Second
_________
6. Duration (
dur
)
0 to 299.9 sec
2 Seconds
_________
D. Temperature Setup
(
t-c
)
1. Display Temperature (
d-t
) (
oC
/
oF
)
°C
_________
2. Automatic (
t-C
) (
Atc
) (
on
/
oFF
)
On
_________
E. Current Output Setup
(
(
OUt
)
1. mA Output (
CUr
) (
020
/
420
)
420
_________
2. Display Current Output (
d-O
) (
Pct
/
doc
)
doc
_________
3. Dampen Current Output (
dPn
)
0-255 sec.
0.0 Seconds
_________
F. Displays Convention
1. U.S. (
Or
)/European (
rO
)
Or
__________
G. Default Setup
(
dFt
)
1. Relay 1 Default (
rL1
) (
non
/
oFF
/
on
)
non
_________
2. Relay 2 Default (
rL2
) (
non
/
oFF
/
on
)
non
_________
3. Current Output Default (
Cur
) (
non
/
cur
)
non
_________
H. Keyboard Security Setup
1. Keyboard Security Required
001-999
_
_________
2. Keyboard Security Not Required
000
000
_________
Alarm Set Points
1. Alarm 1 (
SP1
)
-1500 - +1500 mV
-1500 mV
_________
2. Alarm 2 (
SP2
)
-1500 - +1500 mV
+1500 mV
_________
Current Output
1. Zero (0 or 4 mA)
-1500 - +1500 mV
-1500 mV
_________
2. F.S. (20 mA)
-1500 - +1500 mV
+1500 mV
_________
TABLE 4-2. Configuration Worksheet (ORP)
Use this work sheet to assist in the configuration of the analyzer. Date: ____________________