background image

7

PENBERTHY 

SERIES U FLAT GLASS GAGES 

INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

8.3 Reassembly

If all glass seating surfaces are found to be 
within the tolerances described in the previous 
section, proceed to obtain new glass, gaskets 
and cushions, and proceed to reassemble 
as follows (refer to exploded parts view in 
Section 11 if needed):
1.  Clean threads on cap screws, studs and 

nuts to remove all paint, rust and scale. 
Apply a light coat of oil to the threads.

2.  Lay chamber flat on bench with gasket slots 

facing up. Lay out covers along the bench, 
side by side, with the liquid chamber. Use 
chamber to space covers and line them up 
with vision slots.

3.  Install gasket in place on liquid chamber. 

When using a gasket with a retainer, the 
gasket should be installed with the retainer 
side away from the chamber.

WARNING

Separate installation instructions are supplied 
with replacement glass. All instructions supplied 
with the glass must be followed, as there are 
precautions to be taken when handling gage 
glass. Among the precautions is avoidance of 
bumping or sliding glass against any surface and 
inspection of individual pieces. Failure to follow 
any of the replacement gage glass installation 
instructions could result in glass breakage with 
resulting sudden release of pressure, severe 
personal injury or property damage.

4.  Install rubber band around glass, then 

place glass centered on gasket.

5.  Install cushion centered on glass.
6.  Install cover in place being careful to 

maintain components alignment inside.

7.  Secure the cover to the liquid chamber with 

the corner cap screws. The cap screws 
should only be snugged down enough to 
maintain the cover positioning.

8.  If the gage is a multiple section, complete 

the gasket, glass, cushion and cover 
assembly for the remaining sections per 
instructions 3 through 7 above.

9.  Rotate the liquid chamber 180° on the 

bench so as to expose the glass pocket on 
the opposite side of the liquid chamber.

10. Install the gasket, glass, cushion and cover 

for each section per instructions 3 through 7 
above.

11. Turn the gage on its side on the bench. 

Install the nuts on one end of the studs and 
insert studs into all remaining holes in the 
covers for each section.

12. Install nameplate, washer and nuts to studs 

on the opposite side of the gage. Tighten 
nuts with fingers. Using a torque wrench, 
tighten cap screws and nuts in 5 ft·lb (7 Nm) 
increments, following the sequence in 
Figure 1 until the torque values shown in 
Table 2 are reached.

Note: 

depending on gage size there may be 

less bolting than shown in Figure 1. Start at 
the center and follow 'Z' pattern outward to the 
limit of bolting on a specific gage.
Refer to Section 5 for installation and Section 6 
for operation of liquid level gage when 
returning to service.

9 DISPOSAL AT END OF USEFUL LIFE

Penberthy gages are used in a variety of fluid 
applications. By following the appropriate 
federal and industry regulations, the user 
must determine the extent of preparation 
and treatment the gage must incur before its 
disposal. A Material Safety Data Sheet (MSDS) 
may be required before disposal services 
accept certain components.
Metal, glass and polymers should be recycled 
whenever possible. Refer to order and 
applicable technical data sheets for materials 
of construction.

10 TELEPHONE ASSISTANCE

If you are having difficulty with your liquid level 
gage, contact your local Penberthy distributor. 
You may also contact the factory direct at 
(956) 430-2500 and ask for an applications 
engineer. So that we may assist you more 
effectively, please have as much of the 
following information available as possible 
when you call:
∙ Model #
∙ Name of the company from whom you 

purchased the liquid level gage

∙ Invoice # and date
∙ Process conditions (pressure, flow rates, 

tank shape, etc)

∙ A brief description of the problem
∙ Trouble shooting procedures that failed

If attempts to solve your problem fail, you 
may request to return your gage to the 
factory for intensive testing. You must obtain 
a Return Authorization (R.A.) number from 
Emerson before returning anything. Failure 
to do so will result in the unit being returned 
to you without being tested, freight collect. 

NOTE

Viton

®

 and Teflon

®

 are registered trademarks of the 

Chemours Company.

To obtain an R.A. number, the following 
information (in addition to that above) is 
needed:
∙ Reason for return
∙ Person to contact at your company
∙ 'Ship-to' address

There is a minimum charge for evaluation 
of non-warranty units. You will be contacted 
before any repairs are initiated should the cost 
exceed the minimum charge. If you return a 
unit under warranty, but it is not defective, the 
minimum charge will apply.

Summary of Contents for PENBERTHY Series

Page 1: ...essed or implied shall be made by or on behalf of Emerson unless made in writing and signed by the company s general manager or director of engineering PENBERTHY SERIES U FLAT GLASS GAGES INSTALLATION OPERATION AND MAINTENANCE INSTRUCTIONS valves emerson com 2017 Emerson All rights reserved VCIOM 04042 EN 17 05 1 ABOUT THE MANUAL This manual has been prepared as an aid and guide for personnel invo...

Page 2: ... 4 5 7 6000 psig 41370 kPa at 20 F 29 C to 250 F 121 C FKM Viton w 302 STS retainer 1 3 4 5 7 6000 psig 41370 kPa at 20 F 29 C to 400 F 204 C PTFE Teflon special chamber required 1 3 4 5 7 6000 psig 41370 kPa at 20 F 29 C to 500 F 260 C NOTE Lower temperatures are possible with metallic material variation e g 316 stainless construction PTFE Teflon gaskets cushions usable to 325 F 198 C To determin...

Page 3: ... construction of the liquid level gage are compatible with both the contained media and surrounding atmosphere in the specific application 5 INSTALLATION Installation should only be undertaken by qualified experienced personnel who are familiar with equipment of this type They should have read and understood all of the instructions in this manual The user should refer to applicable technical data ...

Page 4: ...pansion loops in the system Failure to allow for expansion or contraction can result in leaks or glass breakage 5 3 Mirror viewing For added safety a system of indirect viewing by means of mirrors should be installed to protect personnel from the hazards of possible gage failure 5 4 Nut retorquing Nut retorque is vital to the operation of a liquid level gage because gaskets take permanent set unde...

Page 5: ...er and a soft cloth Inspect the surface of the glass for any clouding etching or scratching or physical damage such as bruises checks or corrosion Glass that is damaged is weakened and may break under pressure 7 2 Troubleshooting Problem Glass becomes etched prematurely or clouded in service Cause Fluid being handled is not compatible with the glass Solution Check application to determine actual o...

Page 6: ...ING Once used cushions and gaskets are deformed permanently by compression and if re used may cause leaks and high stress points resulting in glass breakage Glass may contain hidden damage and internal stresses caused by previous usage If re used the glass may break under pressure causing severe personal injury or property damage FIGURE 2 Nut loosening sequence 8 2 Inspection of glass seating surf...

Page 7: ... Install nameplate washer and nuts to studs on the opposite side of the gage Tighten nuts with fingers Using a torque wrench tighten cap screws and nuts in 5 ft lb 7 Nm increments following the sequence in Figure 1 until the torque values shown in Table 2 are reached Note depending on gage size there may be less bolting than shown in Figure 1 Start at the center and follow Z pattern outward to the...

Page 8: ...plate 331 Band RECOMMENDED SPARE PARTS Item Description Min qty 3 Stud 2 per sect 4 Nut 2 per ect 48 Glass 1 100 Screw 1 per sect 7 Gasket 2 8 Cushion 2 FIGURE 3 U Transparent PENBERTHY SERIES U FLAT GLASS GAGES INSTALLATION OPERATION AND MAINTENANCE INSTRUCTIONS NOTE Size 7 shown actual gage may be shorter and require fewer bolting components ...

Reviews: