3.4 Switch Setting Adjustment
Access to switches is achieved by removing control unit from base, unclipping plastic panel from rear of control unit
and sliding PCB from housing. After setting switches, reassemble by reversing the disassembly procedure.
Figure 3 :
There is one set of four switches on the pcb of the MES*L unit, which is labelled SW1. This has four colour coded switches
which can be off to the left (in the diagram above) towards the word OFF on the switch itself, or ON to the right, in the direction
of the arrow towards the numbers 1, 2, 3 and 4.
4.0 Technical Specifications
Main relay output
5 amp SPCO resistive 250V a.c.
Fault relay
5 amp SPCO resistive 250V a.c.
Power supply
110/120v a.c. ± 10%
220/240v a.c. ± 10%
50/60 Hz
Temperature range
-20°C to + 65°C
Housing
IP20 enclosure designed for wall mounting
with 2 screws. Optional DIN rail adaptor
available
Dimensions : mm
H = 150 x W = 75 x D = 112
Mounting holes
2 x 3mm Ø on C/L 100mm apart
Sensor power (for 1 or 2 separate Mobrey
Electrosensor)
12v DC nominal
Intrinsically safe
ATEX
II (1) G [EEx ia] IIC
LED Indicators
1 (Green) main relay on
2 (Yellow) main relay off
3 (Yellow) fault relay off
NOTE: Fault relay is energised in normal state,
de-energised or off in the fault condition
Max. cable resistance each core
100 ohms
Weight
600 grams
Maximum altitude
2000 m
Maximum humidity
95%