6.5.3 Continuous Mode
In this mode, ultrasonics are applied to the sample until you stop the cycle. The user
controls the timing of ultrasonics manually. The power supply starts when a START
condition is received and stops when a STOP condition is received.
The following table shows the parameters of Continuous mode, along the default, max.
and min. values.
Entering an illegal value will generate 3 beeps. The system will not accept out-of-range
parameters. (See
for further details).
Table 6.8
Continuous Mode Parameters
Parameter
Default
Max. Value
Min. Value
Amplitude
50%
100%
10%
Ground Detect
0.05 s
0.99 s
0.00 s
NOTICE
You can return to the ready screen without saving any change by pressing the ESC key
at any time.
Table 6.9
Continuous Mode Operational Sequence
Step
Action
Reference
1
Turn on power, wait for the LCD to show ready screen and
current mode.
2
Press the Mode/Configuration key once to get into mode
selection. Current selected mode icon will blink.
Summary of Contents for Branson LPX
Page 4: ...iv 100 412 196 REV 06 ...
Page 6: ...vi 100 412 196 REV 06 ...
Page 8: ...viii 100 412 196 REV 06 ...
Page 10: ...x 100 412 196 REV 06 Table 7 2 System Trouble Analysis Chart 73 Table 7 3 Alarms Errors 76 ...
Page 24: ...14 100 412 196 REV 06 ...
Page 33: ...100 412 196 REV 06 23 Chapter 3 Delivery and Handling 3 1 Delivery and Handling 24 ...
Page 38: ...28 100 412 196 REV 06 4 1 4 Declaration of Conformity Figure 4 1 Declaration of Conformity ...
Page 76: ...66 100 412 196 REV 06 ...
Page 85: ...100 412 196 REV 06 75 Figure 7 2 LPX Power Supply Interconnect Diagram ...