Specification
Value
Shock
The unit shall survive a 1-meter, all sides and corners drop test onto
concrete.
B.5
Intrinsically Safe electrical parameters
Table 1: Device Communicator communication module
FOUNDATION
™
fieldbus
FOUNDATION
™
fieldbus
HART
®
(non-FISCO)
(FISCO)
FF + and -
FF + and -
HART + and -
Ui
30 Vdc
30 Vdc
30 Vdc
Ii
380 mA
215 mA (IIC) 380 mA (IIB)
200 mA
Pi
1.3 W
1.9 W (IIC) 5.3 W (IIB)
1.0 W
Ci
0
0
0
Li
0
0
0
Uo
1.89 V
1.89 V
1.89 V
Io
32 µA
32 µA
32 µA
Po
61 µW
61 µW
61 µW
Co
14.3 µF
14.3 µF
14.3 µF
Lo
100 mH
100 mH
100 mH
Table 2: Device Communicator Plus communication module
mA
interface
FOUNDATION
™
fieldbus
HART
®
FOUNDATION
™
fieldbus
(non-FISCO)
(FISCO)
mA
FF pwr
and F-
FF + and -
HART +
pwr
HART +
and -
FF pwr and
F -
FF + and -
Ui
30 Vdc
17.5 Vdc
30 Vdc
30 Vdc
30 Vdc
17.5 Vdc
30 Vdc
Ii
200 mA
380 mA
380 mA
200 mA
200 mA
380 mA
215 mA (IIC)
380 mA (IIB)
Pi
1.0 W
1.3 W
1.3 W
1.0 W
1.0 W
1.3 W
1.9 W (IIC)
5.3 W (IIB)
Ci
0
231 nF
0
0
0
231 nF
0
Li
0
0
0
0
0
0
0
Uo
0
17.31 V
1.89 V
25.69 V
1.89 V
17.31 V
1.89 V
Technical specifications
User Guide
187
Summary of Contents for AMS Trex
Page 1: ...User Guide Rev 3 August 2017 AMS Trex Device Communicator User Guide ...
Page 8: ...Contents vi User Guide ...
Page 130: ...Field Communicator application 122 User Guide ...
Page 186: ...Fieldbus Diagnostics application 178 User Guide ...
Page 198: ...Technical specifications 190 User Guide ...
Page 206: ...Glossary 198 User Guide ...