15
8.2 “Set-Point”, Cl (on/off)
Cl reading values can be set to operate the Chlorine pump using 2 set-points into
On/Off mode or Proportional (%).
On/Off mode set the instrument to operate using two set values that enable or disable the Chlorine pump. To
use this mode move cursor on “Working Mode”. Press the wheel and select it.
ON/OFF mode
Set Cl value at 0.80 mg/l ON and 1.00 mg/l OFF. The difference between the two Cl values is called HYSTERESIS.
Instrument will enable the Chlorine pump when reading value will decrease at 0.80mg/l
At 0.80mg/l the Chlorine pump will be enabled until reading value will increase at 1.00mg/l.
Waiting Time: to let pump operate at pulses per minutes add one or more minute (1pulse every xx minutes). Otherwise (00 minutes) pump will
operate at set stroke settings (see page 22).
0.80
1.00
ON
OFF