ESD3000- System
52/66
13.3 Module Calibration
Together with an optional high-
impedance high-voltage divider
(ESD-VERI-V), a recalibration and
adjustment of the discharge-
modules (DMx) can be performed
and stored in the ESD3000 memory.
13.4 Firmware Upload
ATTENTION
Incorrect use of the firmware upload may result in permanent data loss ! See Warning
windows before the upload starts.
How to update the ESD3000 Firmware:
1. Using The download module, connect ESD3000 to
the serial port of a computer
2. Switch on ESD3000. Make sure system power is
maintained during the upload process (Batteries
fully charged before start of update ). This takes
approximately 10 minutes.
3. The window indicates the firmware version on the
ESD3000
and the version on the hard disk.
4. If communication with the upload monitor inside the
ESD3000 is established (message...connected to
ESD3000...), click on ‘Start Update’ and wait until
the message ‘...Upload successful...’ appears.
5. Restart the ESD3000 and check if the new version
is shown in the display.