SD32+
System
General Warnings and Cautions
1.5. electromagnetic compatibility
This device is designed for use in environments that comply with IEC 60601-1-2:2001 as indicated
in the tables shown. You must ensure that the device is used in compliance with the IEC
standards.
Emission Test
Compliance
Electromagnetic Environment
Radiated and Conducted RF
emission
CISPR 11
Class B
The device is suitable for use in domestic and
other environments that are connected to a low
voltage power supply.
Group 1
The internal operation uses a low RF. The
emission is not likely to cause interference with
nearby electronic equipment.
Harmonic emission
IEC 61000-3-2
Complies
The device may be connected directly to a low
voltage supply using an electrical ground.
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Complies
The device may be connected directly to a low
voltage supply using an electrical ground.
Table 1. electromagnetic emission
Immunity Test
IEC 60601-1-2 Test Level
Compliance
Electromagnetic
environment
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
6 kV direct contact
8 kV through air
IEC 60601-1-2 Test
Levels
Residential
(Note 1)
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power supply line
1 kV for input/output line
>3m
IEC 60601-1-2 Test
Levels
Residential
(Note 2)
(Note 3)
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
IEC 60601-1-2 Test
Levels
Residential
(Note 2)
(Note 3)
Voltage dips,
short
interruptions
and voltage
variation on
power supply
input lines
IEC 61000-4-11
0% of rated voltage (voltage
dip 100%) for 0.5 cycles
40% of rated voltage
(voltage dip 60%) for 5
cycles
70% of rated voltage
(voltage dip 30%) for 25
cycles
0% of rated voltage (voltage
dip 100%) for 5 cycles
Residential
(Note 2)
(Note 3)
3