11
Operation
If the equipment – after unpacking – is transported it must be assured that that the equipment is carried
in horizontal position and is hold below the bottom. Before connecting the equipment with the main
source it must be checked that the voltage of the main source is the same than the one stated on the
type label of the equipment (230V~ / 50 Hz). The equipment must only be installed on even areas in
order to allow air to circulate below the equipment. Attention must also be paid that the back side of the
equipment is at least 10 cm away from the wall or other obstacles in order to allow an adequate air
circulation. The power supply cable is to be connected with the equipment and thereafter with the main
source. Switch power supply switch to I (ON) and check whether the green control light is on.
Note: If the equipment is switched on the first time a light smell of heated ceramic fibre may
occur which may last for a short time. This is normal! Switch the time switch clockwise to the
wished time, max. 120 minutes. Adjust the temperature regulator to the wished temperature and
check whether the orange control light is on.
Note: The orange light may jitter a little bit until the adjusted temperature is reached. If the
equipment is switched on at room temperature it will take approx. 30 minutes until the adjusted
temperature is reached. The recommended operating temperature is +150°C to +180°C.
Shall the equipment operate for more than 120 minutes the time regulator is to be adjusted anti clockwise
to the position “ ∞ ”. In order to get a correct sterilisation the insert must not be overloaded. The load must
allow a good circulation of the hot air. When the adjusted temperature is achieved the sterilisation time
shall be between 30 and 120 minutes, depending on legal requirements and / or specifications of the
manufacturers of the items to be sterilised. The max. possible temperature should not be maintained for
a too long time to avoid a damage of the items to be sterilised. Note: The light up of the red light (alarm)
means that the temperature e220°C and that the safety thermostat has been activated. In this case
the equipment shall be switched off and the service shall be informed. When the adjusted sterilisation time
is elapsed the equipment stops automatically.
Insertion / removal of the inserts:
Cleaning
It is recommended to clean the equipment only by using a soft cloth and pure water.
Never use acids or solvents! It is essential that the power is disconnected before cleaning!
Insert to be placed with opening down
Insertion basket to be placed on top with opening up
With the two holders the insert can
easily be removed