2.3.
Quick Start
Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored
automatically and will be recalled when Quick Start is next selected.
QUICK START
V
=
2000V
f
=
5.0kHz
td
=
15.0ms
tr
=
300ms
kop =
L N PE
+/- =
+
T
=
5:00min
START CHANGE
F1
F2
F3
F4
F5
F6
F7
Press
CHANGE
and the test parameters parameter can be changed.
Select the desired parameter with the related function key and change the value by turning the front panel knob.
The cursor allows the user to define the digit to be changed (fast or slow change).
Press
START
and the test starts immediately with the displayed test parameters.
The operator now can navigate with the
Cursor
from parameter to parameter. The blinking parameter can be
changed by turning the front panel knob.
Press
ESC
will bring the user back to the previous menu level.
Page 3
(Change)
QUICKSTART
Voltage
V :
200V
- 4400V
Frequency f
:
0.1kHz
- 1000.0kHz
Duration
td :
0.1ms
- 999.9ms
Repetition
tr :
10ms
- 9999ms
Test time
T :
0:01min - 99:59min
U
f
td
tr
kop
+/-
T
200
5.0
15.0
300
L N PE +
5:00
F1
F2
F3
F4
F5
F6
F7
After starting the test the elapsed testtime is displayed. All
function keys except F2 (manual trigger) can
Stop
the test
routine. After test stop ( Stop, Test OFF), the display keeps for
about 2s.
Burst Spezifikation gemäss IEC 61000-4-4 (2004-07) Ed 2
f =
5kHz
100kHz
tr =
15ms
0.75ms
td =
300ms
300ms