4
care.elzab.p
l
EN
4
5
6
7
Fig. 3
1
replace it with a new filter
carefully insert the internal unit to the enclosure
insert 4 Torx screws
connect the power cord
attach the lower filter cassette
(points 6 - 7 in the lower filter replacement, Fig. 2)
carefully put the device in the proper position
Replacement of the lamp (
Fig.
4)
replacement every 9000 h of operation or 1000 activations
carefully place the device on its side, on a flat surface
remove the lower filter cassette
(points 2 - 3 in the lower filter replacement, Fig. 2)
carefully slide out the whole internal unit from the enclosure
(points 3 - 5 in the upper filter replacement, Fig. 3)
put the internal unit on a flat surface
1
2
3
4
5
hold the lamp at two distant points and turn 90 degrees
to release the lock
carefully remove the lamp
replace the lamp
turn the lamp 90 degrees to lock it
6
7
8
12
11
9
8
10
7