Quick guide for TBA firmware V1.070
VI. Thermal protection.
(figure 44)
The “Temperature Exceeded” information
will appear when the temperature of the chamber exceeds 100°C
or thermal security measures of the heater are activated.
Solution to the problem: Check if the device has proper ventilation.
Cancel the emergency message following instructions in point 10.
If the alarm occurs again, contact the support service.
In order to enable further work, the reason for the alert should be
eliminated, e.g. the Emergency STOP safety switch should be turned on,
the [F3] - [Failures] button should be pressed.
The display will show information about emergencies, e.g. “Emergency
STOP”.
(figure 45)
In case of any of the emergencies, after their cancellation or termination,
press the [F3] - „Failures” button to open the list of failures and cancel
the alert by means of the [F2] - „Cancel” button.
Failures can not be deleted if the emergency is not resolved, e.g. the
emergency switch is on.
Press the [F2] - „Cancel” button to delete the emergency and then
press the [F1] - „Exit” button.
The basic screen can be seen on the display
(figure 46)
.
Press the [F1] - „Exit” button to leave the basic screen and allow for
the further work of the device.
The battery tester is protected from power failure. During the
discharge process current data are saved in a non-volatile memory
every 5 minutes. If power failure and power return occurred during
the test, the display will show the message in
(figure 47)
.
Press the [F3] - [Yes] button to continue the process.
Checking the voltage of the connected battery.
(figure 48)
If the value of the displayed voltage is consistent with the voltage of
the tested battery, press the [F3] - „Next” button.
Then the tester will display the parameters of the battery before the
power outage and will be prepared to continue the test.
To cancel the further discharge process, press the [F1] - „Cancel” button.
figure 45
figure 46
figure 47
figure 44
10. Responding to alerts.
11. Protection from power outage .
figure 48