57
1
Carrier gas inlet
15
Dust filter cartridge
2
Pressure regulator
16
Pressure regulator
3
Oxygen stop valve (power switch
control)
17
Flow adjustment restrictor
4
Pressure switch and pressure gauge
18
Gas flow meter
5
Oxygen stop valve (software
controlled)
19
Infrared cell
6
Flow adjustment restrictor
20
Gas flow meter
7
Carrier gas flow meter
21
Attenuator before the pump
8
Gas pre-cleaning unit
22
Gas flow pump
9
Oxygen diverting valve
23
Attenuator after the pump
10
Flow adjustment restrictor
24
Microcontroller
11
Gas flow meter
25
Computer
12
Upper furnace closure assembly
26
Monitor
13
Flow adjustment restrictor
27
Printer
14
Moisture trap
28
Balance
The analyzer is operated with pure oxygen. A purity of 99.5% is sufficient. It is normally
available in steel bottles. CO2 and H2O impurities in the oxygen are trapped in the trap (8). The
upper half of the trap is filled with CO2 absorber and the lower half with moisture absorber.
Magnesium perchlorate (anhydrone) acts as a H2O absorber.
Sodium hydroxide acts as CO2 absorber, preferably with an indicator, so that the degree of
saturation can be seen from the coloration.
The oxygen inlet pressure should be 2 to 4 bar, which is then regulated inside the analyzer to
1.5bar, as shown on the pressure gauge (4). Any pressure fluctuation of the external oxygen
supply has no influence on the accuracy of the measurements.
A pressure switch (4) becomes conductive at around 1 bar (15 psi). This information is passed
to the software which energizes the valve (5) and starts the gas pump (22) as long as the
pressure is higher than 1 bar. This ensures that the pump runs only when sufficient pressure of
oxygen is available, so that no air enters the gas flow system, if there is no input oxygen
pressure.
With the help of the adjustable restrictor (6) the inlet flow is set at a level of at least
200 l/h. The flow rate is shown on the lower flow meter (7) on the front panel. This flow rate
enters the furnace. During analysis when the furnace is closed, the inlet flow rate is split. One
part enters the furnace top to exit the lance over the crucible. This flow is shown on the middle
flow meter (11). The rest of the oxygen flow enters the furnace at the low end. The ratio of the
two flows can be adjusted with the restrictors (10) and (13).See chapter
The flow at the outlet of the furnace is created by a pump (22) at the end of the gas flow system,
which sucks a flow of 180L/h out of the furnace.
A tube connects the furnace outlet with the moisture trap (14). The gases that come out of the
furnace normally contain some moisture. This particularly happens when analyzing coal
samples. The dried gases pass thru the dust filter (15) and then enter the Infrared cells which
are accommodated in the cell module (19), where the measurements of CO2 and SO2 take
place.
Summary of Contents for CS-580A
Page 1: ...1 Manual Carbon Sulfur Analyzer CS 580A HELIOS REV0004 04 19 Translation ...
Page 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Page 56: ...56 8 2 Gas flow system ...
Page 67: ...67 9 6 TIC module 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Page 70: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...